Correlation between local instantaneous dose rate and oxygen pressure reduction during proton pencil beam scanning
Eleni Kanouta1,2, Jacob Graversen Johansen1,2, Sara Poulsen3
1Danish Centre for Particle Therapy, Aarhus University Hospital, Aarhus, Denmark.
Background And Purpose:
Oxygen dynamics may be important for the tissue-sparing effect observed at ultra-high dose rates (FLASH sparing effect). This study investigated the correlation between local instantaneous dose rate and radiation-induced oxygen pressure reduction during proton pencil beam scanning (PBS) irradiations of a sample and quantified the oxygen consumption g-value.
Materials And Methods:
A 0.2 ml phosphorescent sample (1 μM PtG4 Oxyphor probe in saline) was irradiated with a 244 MeV proton PBS beam. Four irradiations were performed with variations of a PBS spot pattern with 5 × 7 spots. During irradiation, the partial oxygen pressure (pO2) was measured with 4.5 Hz temporal resolution with a phosphorometer (Oxyled) that optically excited the probe and recorded the subsequently emitted light. A calibration was performed to calculate the pO2 level from the measured phosphorescence lifetime. A fiber-coupled scintillator simultaneously measured the instantaneous dose rate in the sample with 50 kHz sampling rate. The oxygen consumption g-value was determined on a spot-by-spot level and using the total pO2 change for full spot pattern irradiation.
Results:
A high correlation was found between the local instantaneous dose rate and pO2 reduction rate, with a correlation coefficient of 0.96-0.99. The g-vales were 0.18 ± 0.01 mmHg/Gy on a spot-by-spot level and 0.17 ± 0.01 mmHg/Gy for full spot pattern irradiation.
Conclusions:
The pO2 reduction rate was directly related to the local instantaneous dose rate per delivered spot in PBS deliveries. The methodology presented here can be applied to irradiation at ultra-high dose rates with modifications in the experimental setup.
More Related Videos
Related Concept Videos
Vapor Pressure Lowering
Measuring Reaction Rates
Rate-Determining Steps
In a multistep reaction mechanism, one of the elementary steps progresses significantly slower than the others. This slowest step is called the rate-limiting step (or rate-determining step). A reaction cannot proceed faster than its slowest step, and hence, the rate-determining step limits the overall reaction rate.