Precision and accuracy of four handheld blood lactate analyzers across low to high exercise intensities
Fredrik Mentzoni1, Martin Skaugen1, Ingrid Eythorsdottir1
1The Norwegian Olympic Sports Center, The Norwegian Olympic and Paralympic Committee and Confederation of Sports, Oslo, Norway.
Purpose:
To evaluate the precision and accuracy in measured blood lactate concentrations among four commonly used handheld lactate analyzers compared to two stationary analyzers.
Methods:
Venous blood samples were taken at exercise intensities ranging from low to high. The blood lactate concentration was measured simultaneously with four pairs of handheld lactate analyzers (two new units of each brand: Lactate Plus, Lactate Pro2, Lactate Scout 4, and TaiDoc TD-4289), and compared with two stationary analyzers (Biosen C-Line and YSI Sport 1500). Measurements were repeated for a range of blood lactate concentrations (measured with Biosen) from 0.88 to 4.89 mM with a median difference between measurements of 0.10 mM.
Results:
The mean relative differences to the Biosen analyzer were 7% (Plus), 7% (Pro), 10% (Scout), 42% (Tai), and 32% (Ysi). The residual standard errors after linear regression against Biosen were 0.18 mM (Plus), 0.20 mM (Pro), 0.22 mM (Scout), 0.15 mM (Tai), and 0.06 mM (Ysi). Accordingly, a blood lactate concentration of 3 mM measured with Biosen yielded 95% prediction intervals that were 0.72 mM (Plus), 0.80 mM (Pro), 0.87 mM (Scout), 0.60 mM (Tai), and 0.23 mM (Ysi) wide.
Conclusion:
Compared to our two stationary analyzers, the precision of the four handheld lactate analyzers evaluated in this study was poor. Among the four, Tai was the most precise; however, this analyzer had low accuracy with a substantial mean difference to the reference analyzer.
More Related Videos
04:28Evaluation of Blood Lactate and Plasma Insulin During High-intensity Exercise by Antecubital Vein Catheterization
Published on: May 18, 2018
07:32Experimental Protocol of a Three-minute, All-out Arm Crank Exercise Test in Spinal-cord Injured and Able-bodied Individuals
Published on: June 8, 2017
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...