MRM Microcoil Performance Calibration and Usage Demonstrated on Medicago truncatula Roots at 22 T
Published on: January 16, 2021
Normalization of increasing shocking coil impedance with full output synchronized shock
Takahide Kadosaka1, Masaya Watanabe1,2, Motoki Nakao1
1Department of Cardiovascular Medicine, Hokkaido University Graduate School of Medicine, Sapporo, Japan.
Elevated low voltage sub-threshold measurement (LVSM) shock impedance (SI) in patients with cardiovascular implantable electronic devices (CIEDs) may be inaccurate. A full output shock test can help assess high voltage shock impedance (HVSI) and normalize LVSM-SI.
Area of Science:
- Cardiovascular implantable electronic devices (CIEDs)
- Electrophysiology
- Biomedical engineering
Background:
- Impedance monitoring is critical for CIED function.
- Low voltage sub-threshold measurement (LVSM) is commonly used for impedance assessment.
- LVSM of shock impedance (LVSM-SI) may become inaccurate due to peri-lead tissue changes.
Observation:
- A case of elevated LVSM-SI was observed 8 years post-lead implantation.
- This elevation was potentially caused by right ventricular lead coil encapsulation.
- High voltage shock impedance (HVSI) measured after a test shock was normal.
Findings:
- A full output synchronized shock test accurately assessed HVSI.
- Post-test shock, LVSM-SI normalized and remained stable during follow-up.
- This suggests a method to correct inaccurate LVSM-SI readings.
Implications:
- Consider a full output synchronized shock test for abnormal LVSM-SI in remote post-ICD implantation phases.
- This approach may help normalize LVSM shock impedance measurements.
- Ensures reliable impedance monitoring for CIEDs.
Related Concept Videos
Design Example: Automobile Ignition System
One can generate a large voltage using a car battery of 12 volts with the help of inductors. Inductors are known for opposing...
Voltage Doubler Circuit
Magnetic Damping
If, however, the bob is a slotted metal plate, the magnet produces a much smaller effect. When a slotted metal plate enters the field, an emf is induced by the change in flux; however, it is less effective because the slots limit the...
Magnetic Field Due to Two Straight Wires
Line Protection with Impedance Relays
Under normal conditions, low load currents keep the measured...
Insulation Coordination