Screening Mammography Diagnostic Reference Level System According to Compressed Breast Thickness: Dubai Health
Entesar Z Dalah1,2, Maryam K Alkaabi3, Hashim M Al-Awadhi3
1HQ Diagnostic Imaging Department, Dubai Health, Dubai 2727, United Arab Emirates.
Journal of Imaging
|August 28, 2024
Summary
This study developed a Diagnostic Reference Level (DRL) system for screening mammography to optimize radiation dose and identify centers needing improvement. The system effectively highlights healthcare centers requiring urgent action for radiation safety.
Area of Science:
- Medical Imaging
- Radiological Physics
- Public Health
Background:
- Screening mammography is crucial for early breast cancer detection.
- Longitudinal exposure may increase radiation-induced breast cancer risk, necessitating dose optimization.
- Current practices require robust systems for internal auditing and dose monitoring.
Purpose of the Study:
- To establish a comprehensive Diagnostic Reference Level (DRL) system for screening mammography.
- To identify healthcare centers with elevated radiation doses requiring optimization.
- To create a tool for dose notification level assessment.
Main Methods:
- Retrospective analysis of screening mammography data from 2048 women across seven centers.
- Establishment of typical DRLs based on bilateral image views (LCC, RCC, LMLO, RMLO) and compressed breast thickness (CBT) criteria.
- Classification of CBT into eight groups (e.g., <20 mm to 90-110 mm) and calculation of local DRLs per view and CBT group.
Main Results:
- A comprehensive DRL system was built using 8877 mammography images.
- Significant dose variations were found among the seven healthcare centers.
- Local DRLs for the 60-69 mm CBT group were 1.24 mGy (LCC), 1.23 mGy (RCC), 1.34 mGy (LMLO), and 1.32 mGy (RMLO).
Conclusions:
- The developed DRL system is efficient, user-friendly, and clinically effective.
- The system successfully identified at least one center needing optimization per bilateral image view and CBT group.
- Implementation of this DRL system supports radiation dose optimization in screening mammography.