Novel In-Line Hemolysis Detection on a Blood Gas Analyzer and Impact on Whole Blood Potassium Results
Shankar Balasubramanian1, Emily J McDowell1, Erving T Laryea2
1Werfen, Research and Development, Bedford, MA, United States.
Background:
Preanalytical error due to hemolyzed blood samples is a common challenge in laboratory and point-of-care (POC) settings. Whole blood potassium (K+) measurements routinely measured on blood gas analyzers are particularly susceptible to hemolysis, which poses a risk for incorrect K+ results. The GEM Premier 7000 with IQM3 (GEM 7000) blood gas analyzer provides novel integrated hemolysis detection within the sample measurement process. Therefore, the GEM 7000 can detect and flag hemolyzed whole blood samples at the POC, warning the operator of potentially erroneous results.
Methods:
Heparinized venous or arterial whole blood samples were used for K+ interference studies and assessed for hemolysis agreement utilizing either a traditional volumetric method or chemistry analyzer serum index measurements with the Roche cobas c311 or Abbott Alinity c.
Results:
Hemolysis interference studies performed at 2 different K+ concentrations (3.8 and 5.3 mmol/L) identified that a plasma free hemoglobin ≥116 mg/dL can impact K+ results on the GEM 7000. Hemolysis agreement studies demonstrated an excellent agreement of >99% with the volumetric method, 98.8% with cobas H index, and 96.4% with Alinity H index. GEM 7000 K+ results were correctly flagged for both native and spiked samples.
Conclusion:
GEM 7000 hemolysis detection provides a novel technology to detect hemolysis in whole blood samples. Moreover, the GEM 7000 demonstrates excellent agreement with traditional laboratory hemolysis detection methods and offers an integrated technological solution for assuring the quality of whole blood K+ results in POC settings.
More Related Videos
05:18A Rapid and Chemical-free Hemoglobin Assay with Photothermal Angular Light Scattering
Published on: December 7, 2016
11:12Ex Vivo Red Blood Cell Hemolysis Assay for the Evaluation of pH-responsive Endosomolytic Agents for Cytosolic Delivery of Biomacromolecular Drugs
Published on: March 9, 2013
Related Concept Videos
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and...