Local and whole-body SAR in UHF body imaging: Implications for SAR matrix compression
Thomas M Fiedler1, Mark E Ladd1,2,3,4, Stephan Orzada1,2
1Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
Purpose:
Transmit arrays for body imaging have characteristics of both volume and local transmit coils. This study evaluates two specific absorption rate (SAR) aspects, local and whole-body SAR, of arrays for body imaging at 7 T and also for a 3 T birdcage.
Methods:
Simulations were performed for six antenna arrays at 7 T and one 3 T birdcage. Local SAR matrices and the whole-body SAR matrix were computed and evaluated with random shims. A set of reduced local SAR matrices was determined by removing all matrices dominated by the whole-body SAR matrix.
Results:
The results indicate that all RF transmit coils for body imaging in this study are constrained by the local SAR limit. The ratio between local and whole-body SAR is nevertheless smaller for arrays with large FOV, as these arrays also expose a larger part of the human body. By using the whole-body SAR matrix, the number of local SAR matrices can be reduced (e.g., 33.3% matrices remained for an 8-channel local array and 89.7% for a 30-channel remote array; 12.1% for the 3 T birdcage).
Conclusion:
For transmit antenna arrays used for body imaging at 7 T as well as for the 3 T birdcage, all evaluated cases show that the local SAR limit was reached before reaching the whole-body SAR limit. Nevertheless, the whole-body SAR matrix can be used to reduce the number of local SAR matrices, which is important to reduce memory and computing time for a virtual observation points (VOP) compression. This step can be included as a pre-compression prior to a VOP compression.