Impact of blood volume, air exposure duration, transport duration, and testing delay on plasma total carbon dioxide
Michael S Reid1, Isolde Seiden Long1, Lawrence de Koning2
1Alberta Precision Laboratories, Diagnostic and Scientific Research Centre, #9 3535 Research Way NW, Calgary, AB T2L 2K8, Canada; Departments of Pathology and Laboratory Medicine, Cumming School of Medicine, Health Sciences Centre, Foothills Campus, University of Calgary, 3030 Hospital Drive NW, Calgary, AB T2N 4N1, Canada.
Background:
Exposing blood specimens to air reduces plasma total carbon dioxide (TCO2). We evaluated the degree of TCO2 reduction attributed to open collection of neonatal blood in BD microtainers® (microtainers), microtainer transport duration and delayed testing of open plasma aliquots.
Methods:
Venous blood was aliquoted into open microtainers in a 3x4 factorial design to simulate combined effects of blood volume (0.2-0.6 mL) and air exposure duration (0-5 min), with blood drawn in vacutainers as a control. Separate effects of in-hospital transport duration (0-120 min; whole blood), off-site transport duration (0-24 h; centrifuged whole blood), and the duration plasma aliquots remained open (0-120 min) were evaluated by repeated testing. Findings were analyzed using repeated-measures ANOVA and Student's T-tests.
Results:
In the factorial experiment, mean plasma TCO2 in microtainers was on average 3.5 mmol/L lower than in vacutainers. Smaller blood volume but not greater air exposure duration significantly (p < 0.05) reduced TCO2. Mean TCO2 in filled (0.6 mL; 1-5 min air exposure) microtainers was on average 2.9 mmol/L lower than in vacutainers. Simulated off-site transport of microtainers containing centrifuged whole blood significantly reduced TCO2 (4 h; mean change = -1.5 mmol/L), as did delayed testing of aliquoted plasma (15 min; mean change = -1.3 mmol/L).
Conclusions:
Plasma TCO2 decreased with reduced microtainer blood volume, extended off-site transport duration of centrifuged whole blood and testing delay of aliquoted plasma. To minimize TCO2 reduction, microtainers should be fully filled and tested rapidly. Laboratories should also consider whether an interpretive comment, correction factor or separate reference intervals are appropriate for these tests.
More Related Videos
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
08:25Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
Related Concept Videos
Carbon Dioxide Transport in the Blood
Forms of CO2 Transport
1. Dissolved in plasma: A small percentage (7-10%) of CO2 is transported and dissolved directly in the plasma.
2. Carbaminohemoglobin: Just over 20% of CO2 is chemically bound to...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Oxygen Transport in the Blood