Craniospinal irradiation using pencil beam Scanning: The PSI experience
Alessandra Bolsi1, Claudio De Angelis1, Miriam Vázquez1
1Center for Proton Therapy, Paul Scherrer Institute, ETH Domain, CH-5232 Villigen, Switzerland.
Summary
Proton craniospinal irradiation (CSI) in children showed no link between dose inaccuracies and local tumor recurrences. Late toxicities were also not correlated with enhanced linear energy transfer (LET), suggesting robust treatment outcomes.
Area of Science:
- Medical Physics
- Radiation Oncology
- Pediatric Oncology
Background:
- Craniospinal irradiation (CSI) is a critical treatment for pediatric central nervous system tumors.
- Proton beam therapy (PBS-PT) offers dosimetric advantages for CSI.
- Evaluating treatment accuracy and outcomes is essential for optimizing pediatric cancer care.
Purpose of the Study:
- To assess the dosimetric accuracy of proton CSI at the Paul Scherrer Institute (PSI).
- To investigate the relationship between treatment accuracy, local recurrences, and late toxicities in pediatric patients.
- To compare treatment outcomes between prone and supine patient positioning.
Main Methods:
- Retrospective analysis of 71 pediatric patients receiving proton CSI between 2004 and 2021.
- Comparison of treatment accuracy between prone (Gantry1) and supine (Gantry2) patient setups.
- Dosimetric evaluation of CSI plans and detailed analysis of local failures and late toxicities in selected patients.
Main Results:
- Supine patient positioning demonstrated improved accuracy with lower systematic (0.25 mm) and random (1.4 mm) errors compared to prone positioning (0.4 mm and 1.8 mm, respectively).
- No correlation was found between local failures and dose inaccuracies or treatment plan robustness.
- No association was observed between patient toxicities and enhanced linear energy transfer (LET).
Conclusions:
- Proton CSI at PSI is dosimetrically accurate and robust, with supine positioning offering improved setup accuracy.
- Treatment accuracy does not appear to influence local recurrence rates in pediatric CSI.
- Observed late toxicities are not linked to enhanced LET, suggesting a favorable safety profile for proton CSI.