Comparison of single shot and multishot diffusion-weighted imaging in 5-T magnetic resonance imaging for brain

Hao Chen1, Runyu Tang2, Xiaopeng Song2

  • 1Department of Radiology, the First Affiliated Hospital of USTC, Division of Life Sciences and Medicine, University of Science and Technology of China, Hefei, China.

Abstract

Insights

Multishot echo-planar imaging (msEPI) offers superior image quality and reduced distortion compared to single-shot echo-planar imaging (ssEPI) for brain disease detection at 5.0 T. This advanced technique shows promise for routine clinical use in diagnosing brain disorders.

Area of Science:

  • Radiology
  • Neuroimaging
  • Magnetic Resonance Imaging (MRI)

Background:

  • Single-shot echo-planar imaging (ssEPI) is crucial for detecting acute brain lesions but suffers from image distortions.
  • Multishot echo-planar imaging (msEPI) is an alternative technique designed to minimize these distortions.
  • High-field MRI at 5.0 T offers enhanced signal but can exacerbate image artifacts.

Purpose of the Study:

  • To compare the image quality and diagnostic performance of ssEPI-DWI and msEPI-DWI at 5.0 T for detecting brain diseases.
  • To evaluate the impact of msEPI on image distortions and diagnostic accuracy compared to ssEPI.
  • To assess the potential of msEPI-DWI at 5.0 T for clinical application in neuroimaging.

Main Methods:

  • Retrospective review of 107 patients with suspected brain diseases who underwent both ssEPI-DWI and msEPI-DWI at 5.0 T.
  • Independent assessment of image quality and distortion by two radiologists.
  • Quantitative analysis of signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and apparent diffusion coefficient (ADC) values.
  • Statistical comparison using Wilcoxon and paired t-tests; diagnostic accuracy assessed using the area under the curve (AUC).

Main Results:

  • msEPI-DWI demonstrated significantly superior image quality and reduced anatomical distortion compared to ssEPI-DWI (P<0.001).
  • No significant differences were found in SNR, CNR, or ADC values between the two techniques (P≥0.601).
  • ADC values from msEPI-DWI and ssEPI-DWI showed strong correlations (r=0.97 for lesions, r=0.91 for normal tissues).
  • Both techniques achieved excellent AUCs for detecting low-grade gliomas (LGGs) (0.934 for ssEPI-DWI, 0.944 for msEPI-DWI).
  • ADC values differed significantly between specific lesions (LGGs, acute cerebral infarction) and normal white matter for both techniques (P≤0.003).

Conclusions:

  • Multishot echo-planar imaging (msEPI) at 5.0 T provides improved image quality and reduced distortion over single-shot echo-planar imaging (ssEPI) for various brain diseases.
  • msEPI-DWI shows promising diagnostic performance for conditions like low-grade gliomas and acute cerebral infarction.
  • The findings suggest that msEPI-DWI at 5.0 T could become a standard clinical tool for diagnosing and grading brain disorders.