Use the external radiotherapy StarTrack detector to study the electron beam profile

Safa Sami Hassan1

  • 1Department of Radiotherapy and Nuclear Medicine, University of Baghdad, Baghdad, Iraq.

Summary

Electron beam dose verification using the StarTrack detector revealed output dose instability for 12MeV energy. The dose exceeded tolerance levels, requiring correction before patient treatment to ensure radiotherapy accuracy.