Application of a Deep Learning-Based Contrast-Boosting Algorithm to Low-Dose Computed Tomography Pulmonary
Minsu Park1, Minhee Hwang2, Ji Won Lee2
1Department of Radiology and Research Institute for Convergence of Biomedical Science and Technology, Pusan National University Yangsan Hospital, Pusan National University School of Medicine, Yangsan, South Korea.
Objective:
The aim of this study was to assess the effectiveness of a deep learning-based image contrast-boosting algorithm by enhancing the image quality of low-dose computed tomography pulmonary angiography at reduced iodine load.
Methods:
This study included 179 patients who underwent low-dose computed tomography pulmonary angiography with a reduced iodine load using 64 mL of a 1:1 mixture of contrast medium from January 1 to June 30, 2023. For single-energy computed tomography, the noise index was set at 15.4 to maintain a CTDI vol of <2 mGy at 80 kVp, and for dual-energy computed tomography, fast kV-switching between 80 and 140 kVp was employed with a fixed tube current of 145 mA. Images were reconstructed by 50% adaptive statistical iterative reconstruction (AR50) and a commercially available deep learning image reconstruction (TrueFidelity) package at a high strength level (TFH). In addition, AR50 images were further processed using a deep learning-based contrast-boosting algorithm (AR50-CB). Quantitative and qualitative image qualities and numbers of involved vessels with thrombus at each pulmonary artery level were compared in the 3 image types using the Friedman test and Wilcoxon signed rank test.
Results:
Five hundred thirty-seven reconstructed image datasets of 179 patients were analyzed. Quantitative image analysis showed AR50-CB (30.8 ± 10.0 and 28.1 ± 9.6, respectively) had significantly higher signal-to-noise ratio and contrast-to-noise ratio values than AR50 (20.2 ± 6.2 and 17.8 ± 6.2, respectively) ( P < 0.001) or TFH (28.3 ± 8.3 and 24.9 ± 8.1, respectively) ( P < 0.001). Qualitative image analysis showed that contrast enhancement and noise scores of AR50-CB were significantly greater than those of AR50 ( P < 0.001) and that AR50-CB enhancement scores were significantly higher than TFH enhancement scores ( P < 0.001). The number of subsegmental pulmonary arteries affected by thrombus detected was significantly greater for AR50-CB (30 for AR50, 30 for TFH, and 55 for AR50-CB, P < 0.001).
Conclusions:
The use of a deep learning-based contrast-boosting algorithm improved image quality in terms of signal-to-noise ratio and contrast-to-noise ratio values and the detection of thrombi in subsegmental pulmonary arteries.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...