Related Experiment Video
Updated: Jun 9, 2025

Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry UPLC-HRMS
Published on: May 20, 2013
Rapid QC-MS: Interactive Dashboard for Synchronous Mass Spectrometry Data Acquisition Quality Control
Wasim Sandhu1, Ira J Gray1, Sarah Lin1
1Chan Zuckerberg Biohub, San Francisco, California 94158, United States.
Rapid QC-MS is an automated dashboard that significantly reduces the time and improves the accuracy of assessing liquid chromatography-coupled tandem mass spectrometry (LC-MS/MS) data quality. This tool helps researchers ensure high-quality untargeted LC-MS/MS data by automating quality control checks.
Area of Science:
- Analytical Chemistry
- Biotechnology
- Data Science
Background:
- High-throughput untargeted liquid chromatography-coupled tandem mass spectrometry (LC-MS/MS) workflows are crucial for biological research.
- Manual quality control (QC) of LC-MS/MS data, including the evaluation of analytical standards, is a significant bottleneck, consuming substantial time and resources.
- Existing methods for real-time data quality assessment are often inefficient and prone to inaccuracies.
Purpose of the Study:
- To develop an automated and interactive dashboard, Rapid QC-MS, for efficient assessment of LC-MS/MS data quality.
- To streamline the process of evaluating analytical controls and identifying low-integrity specimens in high-throughput LC-MS/MS experiments.
- To provide researchers with a user-friendly tool for monitoring instrument performance and data integrity in near real-time.
Main Methods:
- Development of an open-source, interactive dashboard (Rapid QC-MS) using Python libraries and established MS analysis software.
- Integration of automated quality control metrics including instrument sensitivity, retention time shifts, and mass accuracy drift.
- Implementation of local data processing with optional remote syncing (Google Drive) and semiautonomous alerting via messaging applications.
Main Results:
- Rapid QC-MS provides near real-time, browser-viewable quality control results minutes after data file generation.
- The dashboard offers interactive visualizations to detect both acute deviations and gradual drifts in performance metrics over various time scales.
- The system successfully integrates with Thermo Orbitrap workflows, offering a configurable and adaptable solution for diverse LC-MS/MS applications.
Conclusions:
- Rapid QC-MS significantly reduces the time and effort required for manual data curation in untargeted LC-MS/MS workflows.
- The automated quality control system enhances the accuracy and reliability of data integrity assessment.
- Rapid QC-MS provides a fast, straightforward, and open-source solution for collecting high-quality untargeted LC-MS/MS data.
Related Concept Videos
Mass Spectrometers
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview