Atomic Force Microscopy Cantilever-Based Nanoindentation: Mechanical Property Measurements at the Nanoscale in Air and Fluid
Published on: December 2, 2022
Accurate Modelling of AFM Force-Indentation Curves with Blunted Indenters at Small Indentation Depths
Stylianos Vasileios Kontomaris1, Anna Malamou2, Andreas Stylianou1
1Cancer Mechanobiology and Applied Biophysics Group, School of Sciences, European University Cyprus, 2404 Nicosia, Cyprus.
Abstract:
When testing biological samples with atomic force microscopy (AFM) nanoindentation using pyramidal indenters, Sneddon's equation is commonly used for data processing, approximating the indenter as a perfect cone. While more accurate models treat the AFM tip as a blunted cone or pyramid, these are complex and lack a direct relationship between applied force and indentation depth, complicating data analysis. This paper proposes a new equation derived from simple mathematical processes and physics-based criteria. It is accurate for small indentation depths and serves as a viable alternative to complex classical approaches. The proposed equation has been validated for ℎ < 3R (where h is the indentation depth and R is the tip radius) and confirmed through simulations with blunted conical and pyramidal indenters, as well as experiments on prostate cancer cells. It is a reliable method for experiments where the tip radius cannot be ignored, such as in shallow indentations on thin samples to avoid substrate effects.