Deep learning-based whole-brain B1 +-mapping at 7T
Felix Krueger1,2, Christoph Stefan Aigner1, Max Lutz1
1Physikalisch-Technische Bundesanstalt, Berlin, Germany.
Magnetic Resonance in Medicine
|October 27, 2024
Summary
Complex-valued neural networks (NNs) can accurately estimate quantitative B1+ field maps from multi-slice localizer scans at 7T. This accelerates subject-specific calibration for parallel transmission (pTx) MRI, improving imaging efficiency.
Area of Science:
- Magnetic Resonance Imaging (MRI)
- Artificial Intelligence in Medical Imaging
- Radiofrequency (RF) Field Calibration
Background:
- Accurate estimation of the transmit magnetic RF field (B1+) is crucial for quantitative MRI, especially at high field strengths like 7T.
- Subject-specific B1+ calibration is essential for parallel transmission (pTx) techniques to achieve homogenous RF excitation.
- Traditional B1+ mapping methods can be time-consuming, limiting their clinical applicability.
Purpose of the Study:
- To investigate the feasibility of using complex-valued neural networks (NNs) for rapid B1+ map estimation.
- To enable B1+ estimation from multi-slice localizer scans with varying slice orientations in the human head at 7T.
- To accelerate subject-specific B1+ calibration for pTx MRI applications.
Main Methods:
- Acquisition of channel-wise B1+ maps and multi-slice localizers (axial, sagittal, coronal) in 15 healthy subjects at 7T.
- Training of four complex-valued NN configurations using five-fold cross-validation, including one trained on all slice orientations.
- Validation of predicted B1+ maps against reference scans using quantitative metrics and in-vivo testing with dynamic kt-point pulses.
Main Results:
- Predicted B1+ maps showed high similarity to measured maps across different orientations.
- Mean relative errors in magnitude were low: 2.70±2.86% (transversal), 1.82±0.69% (sagittal), and 1.33±0.27% (coronal).
- A network trained on all orientations demonstrated robust B1+ prediction, enabling feasible in-vivo homogenous excitation.
Conclusions:
- Complex-valued NNs are feasible for estimating multi-slice B1+ maps from localizer scans at 7T.
- The proposed method accelerates B1+ calibration, enhancing the efficiency of pTx MRI.