Related Experiment Video
Updated: May 12, 2026

06:40
Insertion of Flexible Neural Probes Using Rigid Stiffeners Attached with Biodissolvable Adhesive
Published on: September 27, 2013
14.7K
Flexible and Extensible Ribbon-Cable Interconnects for Implantable Electrical Neural Interfaces.
Negar Geramifard1, Mahasty Khajehzadeh1, Behnoush Dousti1
1Department of Bioengineering, The University of Texas at Dallas, Richardson, Texas 75080, United States.
ACS Applied Materials & Interfaces
|October 30, 2024
Summary
Researchers developed highly flexible and extensible thin-film ribbon cables for neural implants. A novel lattice design enables up to 300% elongation, improving integration with microelectrode arrays for better neural recording.
Area of Science:
- Biomedical Engineering
- Materials Science
- Neuroscience
Background:
- Implanted neural devices require reliable electrical interconnects.
- Existing interconnects often lack the flexibility and extensibility needed for optimal integration with microelectrode arrays.
- Thin-film technologies offer potential for miniaturized and flexible neural interfaces.
Purpose of the Study:
- To design and characterize novel thin-film ribbon cables for neural stimulation and recording devices.
- To develop flexible and extensible interconnects that integrate seamlessly with thin-film microelectrode arrays (MEAs).
- To investigate a lattice geometry for enhanced cable extensibility without compromising electrical function.
Main Methods:
- Fabrication of multichannel ribbon cables using amorphous silicon carbide (a-SiC), polyimide, and titanium/gold thin films.
- Investigation of linear, serpentine, and open lattice electrical trace designs via photolithography and thin-film processing.
- Characterization of cable flexibility, tensile elongation tolerance, and impedance changes.
- Development and testing of an electrical interconnect process using conductive epoxy for MEA integration.
- In vivo testing of the ribbon cable-MEA system for acute neuronal recording in rat cortex.
Main Results:
- Ribbon cables demonstrated tolerance to flexural bending (50 μm radius) without impedance change.
- Standard designs failed at <5% tensile elongation, while lattice designs achieved 300% elongation without failure.
- Lattice structure's out-of-plane displacement mechanism prevents fracture during high elongation.
- Extensible cables withstood 50,000 cycles of 45% tensile extension without failure.
- Successful acute neuronal recording was achieved using the ribbon cable connected to an a-SiC MEA in rat cortex.
Conclusions:
- Highly flexible and extensible thin-film ribbon cables were successfully designed and fabricated.
- The lattice geometry is crucial for achieving significant elongation tolerance in neural interconnects.
- These cables can accommodate large extensions, aiding surgical implantation and potentially reducing tethering forces on MEAs.
- The developed interconnects show promise for improving chronic neural recording performance and device longevity.