O-cresol Concentration Online Measurement Based On Near Infrared Spectroscopy Via Partial Least Square Regression
Published on: November 8, 2019
Comparative evaluation of accuracy in near-infrared and Raman spectroscopic determinations of component concentration
Haeseong Jeong1, Seung Jee Yang2, Jomjai Peerapattana3
1Department of Chemistry and Research Institute for Convergence of Basic Science, Hanyang University, Seoul, 04763, Republic of Korea.
Background:
Physical characteristics of packed samples, such as packing density, influence subsequent near-infrared (NIR) and Raman spectral features. This potential decrease in accuracy is a concerning issue in practical applications, such as the analysis of pharmaceutical tablets using vibrational spectroscopy. Thus, we compared the accuracy tolerances of both methods under varying packing densities. For evaluation, diffuse reflectance NIR and two different Raman measurements with laser illumination diameters of 1 and 6 mm (referred to as the WAI-1 and WAI-6 schemes, respectively) were used to analyze paracetamol tablets with packing densities of 1.1, 1.17, 1.24, and 1.29 g/cm³.
Results:
In all three measurements, increased packing density resulted in an increase in band intensity and an upward-moving baseline due to the variation in void volume (equivalent to porosity) in the tablets. The Raman spectra acquired using the WAI-6 scheme were less sensitive to packing density variations by averaging out different photon propagations over a large area during spectral acquisition. The accuracies for determining paracetamol concentration using the WAI-6 scheme did not significantly deteriorate when the packing density difference was not large, such as between 1.17 and 1.24 g/cm³ (absolute difference of 0.07 g/cm³). The WAI-6 scheme enabled more robust compositional analysis of the tablets with different packing densities.
Significance:
The WAI-6 Raman scheme provides a more robust compositional analysis of tablets under variations in packing density, making it a preferable choice for practical applications. Using the WAI-6 scheme, an additional effective strategy to minimize accuracy deterioration is further beneficial.
More Related Videos
09:57Ultrafast Time-resolved Near-IR Stimulated Raman Measurements of Functional π-conjugate Systems
Published on: February 10, 2020
09:46Direct Comparison of Hyperspectral Stimulated Raman Scattering and Coherent Anti-Stokes Raman Scattering Microscopy for Chemical Imaging
Published on: April 28, 2022
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...
High-Resolution Mass Spectrometry (HRMS)
¹H NMR Signal Integration: Overview
Attenuated Total Reflectance (ATR) Infrared Spectroscopy: Overview
The ATR process begins by directing a beam...
Infrared (IR) Spectroscopy: Overview
Different compounds display unique properties due to their...