Prosthetic Neck Fracture of a Cementless Titanium Femoral Stem
Zakhia Michael Malhame1,2, Fouad Lamnaouar1, Abdel Massih Abou Chaaya1
1Orthopedics and Traumatology, Centre Hospitalier Victor Dupouy, Argenteuil, FRA.
Abstract:
This case report presents a unique instance of a prosthetic neck fracture in an 81-year-old male with a CERAVER CERAFIT titanium (Ti6Al4V) femoral implant paired with a ceramic-on-ceramic total hip replacement. No fractures of the neck of a titanium femoral stem coated with hydroxy-apatite have been reported in the literature for this type of implant. The patient experienced a sudden, non-traumatic fracture, many years after his primary surgery. The report discusses the clinical presentation, diagnostic findings, known risk factors for femoral neck implant fractures, surgical management, and the outcome.
Related Concept Videos
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...