Computed Tomography-guided Time-domain Diffuse Fluorescence Tomography in Small Animals for Localization of Cancer Biomarkers
Published on: July 17, 2012
Assessment of Photon-Counting Computed Tomography for Quantitative Imaging in Radiation Therapy
Jens Zimmerman1, Gavin Poludniowski2
1Department of Nuclear Medicine and Medical Physics, Karolinska University Hospital, Stockholm, Sweden; Oncology-Pathology, Karolinska Institutet, Stockholm, Sweden.
Purpose:
To test a first-generation clinical photon-counting computed tomography (PCCT) scanner's capabilities to characterize materials in an anthropomorphic head phantom for radiation therapy purposes.
Methods And Materials:
A CIRS 731-HN head-and-neck phantom (CIRS/SunNuclear) was scanned on a NAEOTOM Alpha PCCT and a SOMATOM Definition AS+ with single-energy and dual-energy CT techniques (SECT and DECT, respectively), both scanners manufactured by Siemens (Siemens Healthineers). A method was developed to derive relative electron density (RED) and effective atomic number (EAN) from linear attenuation coefficients (LACs) of virtual mono-energetic images and applied for the PCCT and DECT data. For DECT, Siemens' syngo.via "Rho/Z"-algorithm was also used. Proton stopping-power ratios (SPRs) were calculated based on RED/EAN with the Bethe equation. For SECT, a stoichiometric calibration to SPR was used. Nine materials in the phantom were segmented, excluding border pixels. Distributions and root-mean-square deviations within the material regions were evaluated for LAC, RED/EAN, and SPR, respectively. Two example ray projections were also examined for LAC, SPR, and water-equivalent thickness, for illustrations of a more treatment-like scenario.
Results:
There was a tendency toward narrower distributions for PCCT compared with both DECT methods for the investigated quantities, observed across all materials for RED only. Likewise the scored root-mean-square deviations showed overall superiority for PCCT with a few exceptions: for water-like materials, EAN and SPR were comparable between the modalities; for titanium, the RED and SPR estimates were inferior for PCCT. The PCCT data gave the smallest deviations from theoretic along the more complex example ray profile, whereas the more standard projection showed similar results between the modalities.
Conclusions:
This study shows promising results for tissue characterization in a human-like geometry for radiation therapy purposes using PCCT. The significance of improvements for clinical practice remains to be demonstrated.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET