Automated Kidney Stone Composition Analysis with Photon-Counting Detector CT, a Performance Study-A Phantom Study
Daniel Dillinger1, Stephan Waldeck2, Daniel Overhoff3
1Department of Vascular and Endovascular Surgery, Federal Armed Services Hospital Koblenz, Koblenz, Germany (D.D.); Department of Radiology and Neuroradiology, Federal Armed Services Hospital Koblenz, Koblenz, Germany (D.D., S.W., D.O.).
Background:
For treatment of urolithiasis, the stone composition is of particular interest, as uric acid (UA) stones can be treated by chemolitholysis. In this ex vivo study, we employed an advanced composition analysis approach for urolithiasis utilizing spectral data obtained from a photon-counting detector CT (PCDCT) to differentiate UA and non-UA stones. Our primary objective was to assess the accuracy of this analysis method.
Methods:
A total of 148 urinary stones with a known composition that was measured by the standard reference method infrared spectroscopy (reference) were placed in an abdomen phantom and scanned in the PCDCT. Our objectives were to assess the stone detection rates of PCDCT and the accuracy of the prediction of the stone composition in UA vs non-UA compared to the reference.
Results:
Automated detection recognized 86.5% of all stones, with best detection rate for stones larger > 5 mm in diameter (95.4%, 88.8% for stones larger than 3 mm, 94.7% for stones larger than 4 mm). Depending on the volume, we found a recognition rate of 92.8% for stones larger than 20 mm3 and 94.0% for stones with more than 30 mm3. Prediction of UA composition showed an overall sensitivity and a positive predictive value of 66.7% and a specificity and negative predictive value of 94.5%. Best diagnostic values volume wise were found by only including stones with a larger volume than 30 mm3, there we found a sensitivity of 91.7%, and a specificity of 92.4%. Sensitivity in dependance of the largest diameter was best for stones larger than 5 mm (85.7%), but specificity decreased with increasing diameter (to 91.3%).
Conclusion:
Automated urinary stone composition analysis with PCDCT showed a good automated detection rate of 86.5% up to 95.4% depending on stone diameter. The differentiation between non-UA and UA stones is performed with an NPV of 94.5% and a PPV of 66.7%. The prediction probability of non-UA stones was very good. This means the automatic detection and differentiation algorithm can identify the patients which will not profit from chemolitholysis.
More Related Videos
07:45Estimation of Urinary Nanocrystals in Humans using Calcium Fluorophore Labeling and Nanoparticle Tracking Analysis
Published on: February 9, 2021
11:47Using 2-Photon Microscopy to Quantify the Effects of Chronic Unilateral Ureteral Obstruction on Glomerular Processes
Published on: March 4, 2022
Related Concept Videos
Determination of Crystal Structures
Imaging Studies I: Kidney, Ureter, and Bladder Studies
Imaging Studies II: Ultrasonography
Imaging Studies III: Computed Tomography