Digital Condylar Parameter Assessment Using Cadiax® 2 and Modjaw®.
Smaranda Buduru1, Sara Hafidi1, Oana Almășan1
1Prosthetic Dentistry and Dental Materials Department, Iuliu Hatieganu University of Medicine and Pharmacy, 400006 Cluj-Napoca, Romania.
Dentistry Journal
|November 26, 2024
Summary
Both Modjaw and Cadiax 2 systems show excellent reliability for recording condylar slope and Bennett angle. However, the absolute values measured by each device differ significantly, likely due to technological variations.
Area of Science:
- Dental Mechanics
- Biomedical Engineering
- Craniomandibular Research
Background:
- Assessing the reliability of condylar movement recording systems is crucial for accurate diagnosis and treatment in dentistry.
- Condylar slope and Bennett angle are key parameters in understanding jaw mechanics.
- Two systems, Modjaw and Cadiax 2, are evaluated for their measurement consistency and comparability.
Purpose of the Study:
- To assess the reliability of Modjaw and Cadiax 2 systems in recording condylar slope and Bennett angle.
- To evaluate if measurements are consistent over time (T0 vs. T1) using the same equipment.
- To compare the condylar slope and Bennett angle values obtained from Modjaw versus Cadiax 2.
Main Methods:
- An observational, descriptive, and prospective study was conducted.
- Twenty-five participants (13 female, 12 male, aged 22-27) were included.
- Measurements of condylar slope and Bennett angle were taken at two time points (T0 and T1) using both devices at 3 mm and 5 mm displacements.
Main Results:
- Both Modjaw and Cadiax 2 demonstrated excellent measurement repeatability (Intraclass Correlation Coefficients > 0.90) between T0 and T1.
- Each system showed high reliability when used independently.
- Significant differences were found in the absolute values of condylar slope and Bennett angle measurements between Modjaw and Cadiax 2.
Conclusions:
- While Modjaw and Cadiax 2 are individually reliable for condylar movement recording, their absolute measurements are not interchangeable.
- Technological differences between the Modjaw and Cadiax 2 systems likely explain the discrepancies in recorded values.
- Clinicians should be aware of these device-specific differences when interpreting condylar movement data.