Design and Prototyping of a Novel Triple Lumen Photo-Angioplasty Device: Lumi-Solve-T
Aldous Tria1, Anak Dharma1, Loren Spiegel1
1Monash Institute of Medical Engineering, Monash University, Melbourne, Australia.
Purpose:
A triple lumen iteration of the novel photo-angioplasty drug eluting balloon catheter (DEBc) Lumi-Solve may be compromised by guidewire shadow (GWS)-mediated attenuation of balloon surface drug activation. The current study aimed to design and evaluate a novel triple lumen prototype, designated Lumi-Solve-T, to circumvent these issues.
Methods:
Effects of guidewire shadowing (GWS) on vascular smooth muscle cell (VSMC) proliferation was evaluated using the MTT assay. In-silico modelling of GWS in the novel triple lumen design was conducted. Computer-aided design (CAD) and finite element analysis (FEA) contributed to development of a novel triple lumen catheter. 3D printing of rudimentary and refined prototypes of the catheter together with assembly of a novel fibre-optic (FO) complex and ex-vivo evaluation of the triple lumen device, Lumi-Solve T, was also performed.
Results:
GW insertion in a parallel triple lumen FO: GW port orientation demonstrated significantly reduced inhibition of VSMC proliferation after 7 days confirming the need for an alternative triple lumen design. In-silico analysis identified a multi-fibre FO sleeve design supported uniform, radial and uninterrupted UV365nm light transmission to the angioplasty balloon surface. FEA confirmed a multi-fibre FO ribbon design afforded a practical method of FO sleeve generation and facilitated a novel hub configuration able to afford a FO ribbon to sleeve transition. 3D printed prototypes demonstrated the utility of the novel design.
Conclusions:
A dedicated third port and lumen for the Lumi-Solve FO is required for optimal balloon surface photo-activation. A novel triple lumen design, Lumi-Solve-T, incorporating a ribbon to sleeve FO transition and novel hub design offers a realistic solution to current device limitations.