Intra- and Inter-Rater Reliability, Measurement Error and Criteria and Convergent Validities of the Dualpex Plus for
Daniela Fantin Carro1,2, Leda Tomiko Yamada da Silveira2, Edmund Chada Baracat1,3
1Disciplina de Ginecologia, Departamento de Obstetricia e Ginecologia, Faculdade de Medicina FMUSP, Universidade de Sao Paulo, Sao Paulo, Brazil.
Objective:
To test the Intra- and inter-rater reliability, measurement error and criteria and convergent validities of the Dualpex Plus (DP) for vaginal manometry in women with urinary incontinence (UI).
Design:
This is a clinimetric properties study.
Setting:
University Hospital in Brazil.
Population:
One hundred and two women with UI.
Methods:
Vaginal manometry was performed with DP and Peritron (Pr), in a random order. Intra-rater reliability was tested within a 1-week interval; inter-rater reliability was tested on the same day by two different evaluators. Perception of contraction and comfort with each device were assessed and compared with Chi-square or paired t tests. Reliability was evaluated by intraclass correlation coefficient (ICC) with a 95% confidence interval (CI). Criteria (with Pr) and convergent (with Oxford score) validities were evaluated by Pearson's correlation. The standard error of measurement (SEM) and the minimum detectable change were also calculated.
Main Outcome Measures:
Intra- and inter-rater reliability, measurement error and criteria and convergent validities of the Dualpex Plus.
Results:
DP showed substantial inter-rater and excellent intra-rater reliability (ICC = 0.85; 95% CI 0.76-0.91 and 0.90; 95% CI 0.86-0.93, respectively); strong and positive criteria validity (r = 0.83; p < 0.001) and convergent validity (r = 0.45; p < 0.001). Pressure values for pelvic floor muscle contraction were different between DP and Pr [17.61 ± 12.22 vs. 34.91 ± 21.22 cmH2O; p < 0.001]. SEM was doubtful (19%) and the minimum detectable change was 0.152 cmH2O. DP was more comfortable than Pr and perception of contraction was higher for Pr.
Conclusion:
Dualpex Plus showed adequate clinimetric properties. Measurement error was considered doubtful. DP was more comfortable than Pr and perception of contraction was higher for Pr.
More Related Videos
03:49Anogenital Distance and Perineal Measurements of the Pelvic Organ Prolapse POP Quantification System
Published on: September 20, 2018
07:41Quantification of Levator Ani Hiatus Enlargement by Magnetic Resonance Imaging in Males and Females with Pelvic Organ Prolapse
Published on: April 17, 2019
Related Concept Videos
Reliability and Validity
Muscles of the Pelvic Floor and Perineum
Perineal Layer
The perineum is a diamond-shaped area below the pelvic diaphragm, divided into an anterior urogenital triangle that contains the external genitals and a posterior anal triangle housing the anus. The urogenital...
Assessment of the Rectum and Anus
Rectal Inspection
Begin by inspecting the perianal and anal areas for color, texture, rashes,...
Nursing Assessment of the Genitourinary System II: Inspection and Palpation
Imaging Studies VI: Voiding Cystourethrography and Cystography
Urodynamic Studies: Uroflowmetry