Investigating the Shortcomings of the Flow Convergence Method for Quantification of Mitral Regurgitation in a
Robin Leister1, Roger Karl2,3, Lubov Stroh4
1Institute of Fluid Mechanics (ISTM), Karlsruhe Institute of Technology (KIT), Karlsruhe, Germany. robin.leister@kit.edu.
Abstract:
The flow convergence method includes calculation of the proximal isovelocity surface area (PISA) and is widely used to classify mitral regurgitation (MR) with echocardiography. It constitutes a primary decision factor for determination of treatment and should therefore be a robust quantification method. However, it is known for its tendency to underestimate MR and its dependence on user expertise. The present work systematically compares different pulsatile flow profiles arising from different regurgitation orifices using transesophageal echocardiographic (TEE) probe and particle image velocimetry (PIV) as a reference in an in-vitro environment. It is found that the inter-observer variability using echocardiography is small compared to the systematic underestimation of the regurgitation volume for large orifice areas (up to 52%) where a violation of the flow convergence method assumptions occurs. From a flow perspective, a starting vortex was found as a dominant flow pattern in the regurgant jet for all orifice shapes and sizes. A series of simplified computational fluid dynamics (CFD) simulations indicate that selecting a suboptimal aliasing velocity during echocardiography measurements might be a primary source of potential underestimation in MR characterization via the PISA-based method, reaching up to 40%. In this study, it has been noted in clinical observations that physicians often select an aliasing velocity higher than necessary for optimal estimation in diagnostic procedures.
More Related Videos
11:04Quantification of Global Diastolic Function by Kinematic Modeling-based Analysis of Transmitral Flow via the Parametrized Diastolic Filling Formalism
Published on: September 1, 2014
11:16In vitro Assessment of Aortic Regurgitation Using Four-Dimensional Flow Magnetic Resonance Imaging
Published on: February 25, 2022
Related Concept Videos
Steady Flow of a Fluid Stream
During this process, the momentum of the fluid within the control volume remains constant over the time interval dt. By applying the...
Control Volume and System Representations
The control volume approach considers a stationary region in space through which fluid flows. This region is bounded by a control surface. For instance, in the case of water flowing...
Reynolds Transport Theorem
Conservation of Mass in Finite Cotrol Volume
A system is defined as a collection of unchanging contents, and the conservation of mass states that a system's mass is constant.
Conservation of Mass in Fixed, Nondeforming Control Volume
In the case of a sewer pipe, which can be modeled...
Typical Model Studies