Related Experiment Video
Updated: Jun 2, 2025

05:42
Dynamic Navigation for Dental Implant Placement
Published on: September 13, 2022
3.6K
Implant position accuracy using dynamic computer-assisted implant surgery (CAIS) combined with augmented reality:
Sirida Arunjaroensuk1,2, Thanida Yotpibulwong2, Po-Sung Fu3,4
1Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Chulalongkorn University, Bangkok, Thailand.
Journal of Dental Sciences
|January 14, 2025
Summary
Augmented reality (AR) glasses did not significantly improve the accuracy of dental implant placement when used with dynamic computer-assisted implant surgery (CAIS). This study found similar precision for implant positioning with or without AR technology.
Area of Science:
- Dental Implantology
- Surgical Navigation Technology
- Augmented Reality Applications
Background:
- Computer-assisted implant surgery (CAIS) enhances precision in dental implant placement.
- Dynamic CAIS systems offer real-time visualization of surgical instruments and patient positioning.
- Augmented reality (AR) technology presents potential for overlaying digital information in surgical fields.
Purpose of the Study:
- To compare the accuracy of dental implant positioning using dynamic CAIS with and without AR glasses.
- To evaluate the clinical efficacy of AR glasses in augmenting dynamic CAIS for implant surgery.
Main Methods:
- A randomized clinical study involving 20 patients requiring single dental implants.
- Patients were divided into two groups: dynamic CAIS with AR glasses (n=10) and dynamic CAIS without AR glasses (n=10).
- Three-dimensional deviations (platform, apex, angular) were measured and statistically analyzed.
Main Results:
- The AR glasses group showed mean 3D angular deviations of 1.47° ± 1.01°, compared to 2.42° ± 0.76° in the non-AR group.
- Mean 3D entry point deviations were 0.75 ± 0.45 mm (AR) vs. 1.11 ± 0.44 mm (non-AR).
- Mean 3D apical deviations were 0.87 ± 0.45 mm (AR) vs. 1.18 ± 0.50 mm (non-AR), with no statistically significant differences between groups.
Conclusions:
- Dynamic CAIS with AR glasses demonstrated similar accuracy in single dental implant placement compared to dynamic CAIS alone.
- The integration of AR glasses did not yield a statistically significant improvement in implant position accuracy in this study.
- Further research may explore specific scenarios or advanced AR features for potential benefits in CAIS.
Keywords:
AccuracyAugmented reality (AR)Computer-assisted implant surgery (CAIS)Dynamic systemNavigation