Veno-Venous Extracorporeal Membrane Oxygenation in a Mouse
Published on: October 24, 2018
Impact of Cannulation Strategy and Extracorporeal Blood Flow on Recirculation During Veno-Venous Extracorporeal
Jan-Steffen Pooth1, Jil Kristin Förster2,3, Christoph Benk2
1Department of Emergency Medicine, Faculty of Medicine, University Medical Centre Freiburg, University of Freiburg, Freiburg, Germany.
Introduction:
Veno-venous extracorporeal membrane oxygenation (V-V ECMO) is increasingly used in the treatment of severe respiratory failure. Despite a significant increase in the worldwide use of extracorporeal lung assist devices recirculation remains a common complication and is associated with a reduced effectiveness of ECMO support and increased hemolysis. In this observational study we aimed to investigate the impact of cannula configuration and extracorporeal flow on recirculation.
Materials And Methods:
An observational retrospective study was performed, which included all patients, who received V-V ECMO and recirculation measurements at the University Medical Center Freiburg between August 2021 and June 2023. Recirculation and extracorporeal flow were determined using ultrasonic indicator dilution technology. Patients were divided into subgroups according to their type of cannulation (dual lumen single-site vs. bifemoral vs. femoro-jugular).
Results:
A total of 215 recirculation measurements in 47 patients were performed. Dual lumen single-site cannulation was associated with significantly lower recirculation rates (8.7% [0.0; 12.0]) compared to single lumen dual-site cannulation (femoro-jugular: 17.6% [0.0; 25.8]; bifemoral: 27.9% ± 13.4%). In addition, a positive linear correlation was observed between extracorporeal flow and recirculation in all subgroups. Recirculation increased significantly with rising extracorporeal flow in all subgroups.
Conclusion:
Recirculation is a common complication in V-V ECMO and can lead to a reduction of ECMO effectiveness. Particular attention should be paid to optimal positioning of the cannulas in patients with more than one cannula. The ultrasonic indicator dilution method is a simple and quick method for measuring recirculation in V-V ECMO and can be used at an early stage if effectiveness decreases.
More Related Videos
09:54Lung Rapid Recovery Procurement Combined with Abdominal Normothermic Regional Perfusion in Controlled Donation after Circulatory Death
Published on: August 15, 2022
03:40Point-of-Care Ultrasound for Peripheral Veno-Arterial Extracorporeal Membrane Oxygenation Without Left Ventricular Venting
Published on: January 17, 2025
Related Concept Videos
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy