New Bipolar Electrosurgical Vessel Sealing Device Provides Improved Performance and Procedural Efficiency.
Rheagan Chambers1, Danielle Sarno1, Susan Roweton1
1Surgical Research and Development, Surgical Operating Unit, Medtronic, Lafayette, CO, USA.
The new LigaSure™ XP Maryland Jaw Sealer/Divider (XP Maryland) offers faster sealing times and fewer incomplete cuts compared to earlier LigaSure™ devices. It demonstrates reliable performance, potentially improving surgical efficiency and reducing risks.
Area of Science:
- Surgical Technology
- Medical Devices
- Vascular Surgery
Background:
- LigaSure™ devices are utilized for sealing blood vessels during surgical procedures.
- Early LigaSure™ devices, such as Atlas and Dolphin Tip, have established records of clinical performance.
- Advancements in surgical device technology aim to improve efficiency and safety.
Purpose of the Study:
- To compare the performance of the new LigaSure™ XP Maryland Jaw Sealer/Divider (XP Maryland) against two earlier LigaSure™ devices: LigaSure Atlas™ (Atlas) and LigaSure™ Dolphin Tip (Dolphin Tip).
- To evaluate specific performance metrics including seal times, burst pressures, and rates of procedural complications.
Main Methods:
- Ex vivo bench testing on porcine renal arteries assessing burst pressures, seal times, sticking, incomplete cuts, and charring.
- In vivo acute testing on a porcine model evaluating thermal spread, seal times, hemostasis, sticking, and incomplete cuts.
- Comparative analysis between the XP Maryland and the Atlas and Dolphin Tip devices.
Main Results:
- Ex vivo and in vivo studies showed significantly faster mean seal times for XP Maryland compared to both Atlas and Dolphin Tip.
- XP Maryland demonstrated fewer incomplete cuts and sticking incidents compared to the earlier devices in ex vivo and in vivo tests.
- Atlas exhibited a statistically higher average burst pressure than XP Maryland ex vivo; XP Maryland showed less lateral thermal spread in vivo.
Conclusions:
- The XP Maryland device exhibits consistent and reliable performance, comparable to earlier LigaSure™ models.
- XP Maryland presents improved performance characteristics, including faster sealing and fewer complications, potentially enhancing procedural efficiency and surgical safety.