A validated method for capillary phosphatidylethanol 16:0/18:1 quantification with two different 10-µl volumetric
Trine N Andreassen1, Marina Gule1, Hilde Havnen1
1Department of Clinical Pharmacology, St. Olavs University Hospital, Trondheim 7006, Norway.
Quantifying phosphatidylethanol (PEth) in capillary blood using volumetric absorptive microsampling (VAMS) devices is feasible. Both Mitra and Capitainer devices showed suitability for PEth analysis, with Mitra meeting all validation criteria.
Area of Science:
- Clinical Chemistry
- Toxicology
- Biomarker Analysis
Background:
- Phosphatidylethanol (PEth) is a reliable biomarker for alcohol consumption, traditionally measured in venous whole blood.
- Growing interest exists in quantifying drugs and biomarkers in capillary blood for less invasive sampling.
- Volumetric absorptive microsampling (VAMS) devices offer a promising approach for capillary blood collection.
Purpose of the Study:
- To adapt and validate a laboratory method for quantifying PEth in capillary dried blood spot (DBS) samples.
- To evaluate the performance of two VAMS devices, Capitainer®B Vanadate and Mitra®, for PEth analysis.
- To compare PEth quantification results from capillary DBS samples with established venous blood measurements.
Main Methods:
- Method development and validation for PEth quantification in capillary DBS using two VAMS devices (Capitainer and Mitra).
- Automatic sample extraction was employed, allowing simultaneous analysis of both VAMS types.
- Validation included assessing performance across a defined measuring range and comparing results with a validated venous blood method.
Main Results:
- The Mitra device met all pre-established validation criteria for PEth quantification in the range of 0.03 to 4.0 µM.
- The Capitainer device met most criteria but showed deviations in accuracy for PEth levels ≥ 0.5 µM (-17.1% to -20.5%).
- Correlation analysis indicated an acceptable proportional mean difference (-7.6%) between Capitainer and venous blood results, with no constant bias.
Conclusions:
- Both Capitainer and Mitra VAMS devices are suitable for PEth analysis in capillary blood.
- The Mitra device demonstrated full compliance with validation criteria, making it highly reliable for capillary PEth quantification.
- The Capitainer device showed acceptable performance, though accuracy at higher concentrations warrants consideration for specific applications.
More Related Videos
07:58Highly Sensitive and Quantitative Detection of Proteins and Their Isoforms by Capillary Isoelectric Focusing Method
Published on: September 19, 2018
09:22Absolute Quantitation of Inositol Pyrophosphates by Capillary Electrophoresis Electrospray Ionization Mass Spectrometry
Published on: August 13, 2021
Related Concept Videos
Electrophoresis: Overview
There...
Capillary Electrophoresis: Instrumentation