Background:
The need for transvenous lead extraction (TLE) is increasing worldwide. Since the course of the subclavian vein until the junction with the superior vena cava is a frequent place of lead adherences, we aim to assess the safety and efficacy of a short 13Fr bidirectional rotational mechanical sheath as first choice.
Methods And Results:
In this multicentric study, 202 carriers of a cardiac implantable electronic device (CIED) undergoing TLE using a short 13 Fr bidirectional rotational mechanical sheath were prospectively enrolled. All procedures were performed using a stepwise approach. The indication for TLE were infection (62%), malfunction (32%), and upgrade (6%). Overall, 471 leads were extracted: 65% pacing leads, 20% defibrillator leads, 9% coronary sinus leads, and 6% abandoned leads. Clinical success and complete lead extraction have been achieved in 97% and 95% of cases, respectively. The short sheath was always effective in gaining venous access at the start of the procedure and was sufficient for complete TLE in 67% of cases. Lead dwell time, defibrillator lead, number of leads per patient, and lead malfunction were predicting factors of long bidirectional rotational mechanical sheath use. There were no cases of intraprocedural death, and major complications were reported in 2% of patients. Overall survival was 97% at 1-year follow-up.
Conclusion:
This multicentric experience using a short 13Fr bidirectional rotational mechanical sheath reported high safety and efficacy, demonstrating the utility of routine use of short extraction sheaths of higher caliper as the first choice.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
03:13Author Spotlight: A Single-Entry Point Endoscopic Intraventricular Approach for Third Ventriculostomy and Pineal Biopsy
Published on: June 28, 2024
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization