Automated Injectors versus Manual Administration: A Comparative Analysis of Radiation Exposure Reduction in 18 F-FDG
Saumya Srivastava1, Subhash Chand Kheruka1,2, Sanjay Gambhir1
1Department of Nuclear Medicine, Sanjay Gandhi Postgraduate Institute of Medical Sciences, Lucknow, Uttar Pradesh, India.
Automated injectors significantly reduce radiation exposure for healthcare workers during 18F-FDG administration. This technology is crucial for enhancing safety in nuclear medicine and diagnostic radiology.
Area of Science:
- Nuclear Medicine
- Radiology
- Occupational Health
Background:
- Manual handling of radiopharmaceuticals poses significant occupational radiation risks to healthcare professionals, especially to hands and eyes.
- Existing safety protocols have not fully mitigated radiation exposure during radiopharmaceutical manipulation.
- Nuclear medicine personnel face continuous radiation exposure risks from daily tasks.
Purpose of the Study:
- To evaluate the effectiveness of automated injectors in minimizing radiation exposure for healthcare workers.
- To compare radiation doses from automated versus manual injection of fluorine-18 fluorodeoxyglucose (18F-FDG).
- To assess the impact of automation on whole-body, extremity, and eye-lens radiation doses.
Main Methods:
- Radiation exposure was measured using specialized dosimeters for both manual and automated 18F-FDG injection techniques.
- Dosimetry focused on whole-body, extremity (hands), and eye-lens radiation doses.
- Comparative analysis of radiation levels between the two injection methods was performed.
Main Results:
- Automated injectors reduced left-hand and right-hand extremity doses by 97.97% and 98.96%, respectively.
- Eye-lens radiation dose decreased by 43.24% with automated injection.
- Whole-body radiation dose was reduced by 91.66% when using automated injection systems compared to manual methods.
Conclusions:
- Automated injection systems substantially decrease occupational radiation exposure for healthcare workers in nuclear medicine.
- The significant reduction in staff radiation doses highlights the need to adopt automated technology.
- Implementing automated injectors enhances safety and promotes a safer clinical environment in diagnostic radiology.