Performance of Unmodified Mechanical Ventilators With 2% Hydrogen Gas Mixtures
Julia Garcia Mancebo1, Kristen Sack2, Richard Nguyen3
1Dr. Mancebo is affiliated with Department of Cardiology, Boston Children's Hospital, Boston, Massachusetts; Division of Critical Care, Texas Children's Hospital, Houston, Texas; and Department of Pediatrics, Baylor College of Medicine, Houston, Texas.
Molecular hydrogen (H2) shows therapeutic potential for critical illness. Unmodified ventilators can deliver H2 accurately, except for the Dräger Babylog VN500, which has significant tidal volume inaccuracies.
Area of Science:
- Critical Care Medicine
- Respiratory Therapy
- Biomedical Engineering
Background:
- Molecular hydrogen (H2) possesses antioxidant, anti-inflammatory, and anti-apoptotic properties.
- H2 may benefit ischemia-reperfusion injury and critical illnesses.
- Clinical application of H2 via mechanical ventilators requires performance evaluation.
Purpose of the Study:
- To assess the performance of unmodified mechanical ventilators when delivering 2% H2 mixtures.
- To evaluate the accuracy of tidal volume (VT) and oxygen concentration (FIO2) delivery.
- To identify potential malfunctions or inaccuracies associated with H2 administration.
Main Methods:
- Four ventilator models (Maquet Servo-i, Servo-u, Dräger Evita Infinity V500, Babylog VN500) were tested.
- Ventilators were supplied with certified 2% H2 in air/O2 mixtures.
- Measurements included delivered VT, FIO2, H2 concentration, and pre-use check outcomes across neonatal to adult settings.
Main Results:
- Maquet Servo-i, Servo-u, and Dräger Evita V500 showed acceptable accuracy (within ±10% bias) for VT and FIO2 delivery with H2.
- The Dräger Babylog VN500 exhibited significant inaccuracy in VT delivery (average bias of -89.2%) when using H2 mixtures.
- All tested ventilators, except the Babylog VN500, passed pre-use checks.
Conclusions:
- Unmodified Maquet Servo-i, Servo-u, and Dräger Evita V500 ventilators are suitable for delivering 2% H2 mixtures.
- The Dräger Babylog VN500 is unsuitable for H2 administration due to inaccurate VT delivery.
- Ventilator technology (e.g., hot-wire anemometry) impacts H2 delivery accuracy.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...