[Method of linear calibration function control according to GOST R ISO 11095-2007 for measuring the mass
M S Zatyamina1,2,3, L A Smirnova1,2, O V Ostrovskij1
1Volgograd State Medical University, Volgograd, Russia.
Objective:
Implementation of the control method to validate previously obtained calibration functions according to GOST R ISO 11095-2007 using Reference Materials (RM): standard RMs - aqueous solutions of ethanol, produced and certified by the organization, and internal RMs - aqueous solutions of ethanol in different concentrations, developed by the user for internal use.
Material And Methods:
We applied the control method in accordance with the National Standard of the Russian Federation GOST R ISO 11095-2007 using the standard and internal RMs - aqueous ethanol solutions in different concentrations. Measurements were performed for two subranges of ethanol concentrations on standard and internal RMs in accordance with the certified method.
Results And Conclusion:
Control charts based on experimental data confirmed that the system was in a controlled state, and the calibration function required no updates on Day 7. No studied values fell outside the upper and lower control limits on the control charts. The uncertainty of the transformed values was assessed during the calibration function validation.
More Related Videos
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Flame Photometry: Lab
Constant Volume Calorimetry
UV–Vis Spectroscopy: Beer–Lambert Law