Comparison between Different Configurations of Reference Electrodes for an Extended-Gate Field-Effect Transistor pH
Jiunn-Tyng Yeh1,2, Chen-Yun Hsiao1, Hsuan-Wo Lee1
1Novascope Diagnostics Inc, 2F.-9, No. 38, Taiyuan St., Zhubei 302082, Hsinchu County, Taiwan.
Abstract:
This study investigates the performance and stability of extended-gate field-effect transistor (EGFET) pH sensors with three different reference electrode configurations: conventional external Ag/AgCl, external gold (Au), and integrated Au electrodes. The goal is to identify an optimal reference electrode configuration for enhancing the portability and manufacturability of EGFET pH sensors, which are crucial in medical, agricultural, and environmental applications. Results showed that the integrated Au electrode exhibits superior stability and comparable sensitivity to those of Ag/AgCl and external Au electrodes. The integrated Au electrode configuration achieved a 96 mV/pH sensitivity with 97% linearity across a pH range of 4 to 10. Stability tests revealed that the integrated Au electrode had the least drift under different pH conditions compared with the other two configurations, suggesting improved reliability. These findings highlight the integrated Au electrode's potential as a viable alternative to conventional reference electrodes in EGFET pH sensors, offering enhanced performance and practicality for point-of-care testing and other applications.
More Related Videos
Related Concept Videos
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Potentiometry: Membrane Electrodes
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...
Potentiometry: Overview