A Novel Application of Musculoskeletal Ultrasound Imaging
Published on: September 17, 2013
Reliability of spectral Doppler in handheld ultrasonographic device
Dominika Filipiak-Strzecka1, Jarosław D Kasprzak2, Piotr Lipiec2
1Department and Chair of Cardiology, Bieganski Specialty Hospital, Medical University of Lodz, Kniaziewicza 1/5, Lodz, 91-347, Poland. dominika.filipiak@gmail.com.
Insights
Handheld ultrasound devices (HUD) provide reliable measurements for assessing diastolic function and valvular flow velocities, showing high correlations and good agreement with standard echocardiography.
Area of Science:
- Cardiology
- Medical Imaging
- Ultrasound Technology
Background:
- Assessing left ventricular diastolic function and valvular hemodynamics is crucial for diagnosing heart conditions.
- Traditional echocardiography requires specialized equipment and trained personnel, limiting its accessibility in certain settings.
- Handheld ultrasound devices (HUD) offer a portable alternative for cardiac assessments.
Purpose of the Study:
- To validate the accuracy and reliability of diastolic function parameters and valvular flow velocities obtained using HUD.
- To compare HUD measurements against standard full echocardiographic examinations.
Main Methods:
- A prospective study involving 105 patients referred for dyspnea or exercise intolerance.
- Bedside echocardiographic screening using HUD followed by a reference standard full echocardiogram within 24 hours.
- Analysis included correlation, Bland-Altman plots, and Kappa coefficients for diastolic parameters and valvular velocities.
Main Results:
- High to very high correlations were observed between HUD and standard echocardiography for most measurements.
- Mitral inflow velocities showed a slight underestimation bias, while mitral annular early diastolic velocity (e') had no bias.
- Substantial to almost perfect agreement was found for major HFA-PEFF score parameters; good agreement for accelerated mitral and aortic valve flow detection (Kappa 0.83 and 0.74, respectively).
Conclusions:
- HUD with Doppler capabilities enable reliable assessment of left ventricle diastolic function parameters.
- Accurate measurements of aortic and mitral peak jet velocities can be obtained using HUD.
- HUD can be a valuable tool for rapid and reliable cardiac assessments, particularly in point-of-care settings.
Abstract:
To verify and validate the reliability of diastolic function parameters and valvular flow velocities acquired during the handheld ultrasound device (HUD) echocardiographic examination. Study population consisted of the consecutive patients referred for consultation due to dyspnea or impaired exercise tolerance. All patients underwent brief bedside echocardiographic screening with HUD. Within the next 24 h all patients underwent full echocardiographic examination (treated as reference). 105 patients (58 men, mean age 65 ± 14 years) were enrolled in the study. All correlations of HUD and standard echo derived measurements were high or very high. Bland-Altman plot analysis revealed the underestimation bias for mitral inflow velocities- early (E) and late (A), ), mitral annular peak early diastolic velocity (e') was not burdened with bias. The agreement between the major HFA-PEFF score parameters showed either substantial or almost perfect agreement, minor parameter- moderate agreement. The correlation of the measurements of maximum mitral velocity was high with no clinically relevant bias. Kappa coefficient showed very good agreement between HUD and reference echocardiography for detecting accelerated blood flow through the mitral valve (Kappa coefficient 0.83). The assessment of the aortic valve showed a very good correlation, with a slight underestimation bias of -0.17 m/sec (P <.001). Kappa coefficient showed good agreement between HUD and reference echocardiography for detecting accelerated blood flow (Kappa coefficient 0.74). HUD equipped with pulse and continuous wave Doppler modality enables reliable measurements of the parameters used during the left ventricle diastolic function assessment. Similarly, aortic and mitral peak jet velocities can be accurately measured using the mentioned devices.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Doppler Effect - II
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
UV–Vis Spectrometers