Method for visualizing detailed profiles of synchrotron X-ray beams using diamond-thin films and silicon drift
Togo Kudo1, Shinji Suzuki1, Mutsumi Sano1
1Japan Synchrotron Radiation Research Institute, Sayo-cho 1-1-1, Sayo-gun, Hyogo 679-5198, Japan.
Abstract:
Contamination from nearby bending magnet radiation hinders precise and accurate determination of the true beam center of undulator radiation. To solve this problem, a semi-nondestructive method was developed to visualize the detailed profile of a synchrotron radiation beam by using a thin diamond film as a scatterer. As the beam passed through the diamond film, scattered X-rays were imaged using a pinhole camera and measured with a two-dimensional silicon drift detector (SDD) scan. With this configuration, the beam center was accurately determined by visualizing the radiation pattern distribution for each energy level of a pink X-ray beam within an aperture size of 1.5 mm × 1.5 mm, shaped by a front-end slit (FES) positioned upstream of the monochromator. Additionally, by scanning the FES in two dimensions with a reduced aperture of 0.4 mm × 0.4 mm, energy-resolved images were successfully obtained using the SDD at a fixed position. These images revealed the profile of undulator radiation over a broad area (with an aperture extending up to 4 mm) in a pre-slit positioned upstream of the FES, demonstrating good alignment with SPECTRA calculations. This method effectively eliminates contamination from nearby bending magnet radiation, a significant issue in previous approaches, enabling a direct and highly accurate determination of the true beam center.