Related Experiment Video
Updated: May 10, 2025

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Design and validation of a minibeam treatment delivery system for use with a radiation therapy research platform
Steven Herchko1, Sridhar Yaddanapudi1, C-K Chris Wang2
1Radiation Oncology, Mayo Clinic in Florida, 4500 San Pablo Road, Jacksonville, Florida, 32224-1865, UNITED STATES.
Abstract:
Minibeam radiation therapy (MBRT) is a promising treatment technique in the early stages of clinical use. In this work, two hexagonal minibeam collimators were designed, fabricated, and validated using film measurements and Monte Carlo simulations. Film and Monte Carlo results demonstrated strong agreement, with the greatest agreement near the beam central axis. One of the previously validated MBRT collimators was then modified to allow for dose calculations on a mouse cone beam computed tomography (CBCT) dataset, and a MBRT dose distribution was preserved in the animal dataset. This validated system can be used in future cell and small animal studies to further explore MBRT in a preclinical setting.

