Fetal dose in pregnant CT patients: a comparison of four software packages
Joël Greffier1, Asma Arjoun2, Chris Serrand3
1IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, Nîmes, France. joel.greffier@chu-nimes.fr.
Objectives:
To compare the fetal dose (FD) as calculated by four different software packages for pregnant women who have undergone CT acquisitions directly exposing the whole fetus to X-rays.
Materials And Methods:
Pregnant women who underwent CT abdomen-pelvis and/or thorax-abdomen-pelvis acquisitions from February 2018 to May 2024 and for whom the uterine dose and/or FD was calculated by a medical physicist were retrospectively included. FDs were computed per CT acquisition with VirtualDose-CT™ (VDCT), Duke Organ Dose (DOD), fetaldose.org, and COnceptus Dose Estimation (CODE) software, using phantoms taking the stage of pregnancy into account. FDs calculated by each software package were then compared.
Results:
A total of 51 pregnant women with a mean age of 30.2 ± 5.7 years at 17.5 ± 10.0 weeks of pregnancy were included. The mean number of CT acquisitions per pregnant patient was 1.4 ± 0.7 with a mean CTDIvol of 6.77 ± 3.04 [2.34-15.64] mGy, and FDs were computed for a total of 69 acquisitions. For all CT acquisitions, the median FD was 8.6 (6.8; 10.3) mGy for VDCT, 7.7 (6.1; 9.7) mGy for DOD, 6.3 (4.9; 7.6) mGy for fetaldose.org, and 7.1 (4.6; 8.8) mGy for CODE. Differences between each software package were significant (p < 0.01), except between VDCT and DOD (p = 0.025) and between CODE and fetaldose.org (p = 0.15). The concordance of calculated FD values between the software packages was poor (ICC < 0.50), except between VDCT and CODE and between fetaldose.org and CODE.
Conclusion:
The choice of software used affects the calculation of the FD.
Key Points:
Question Differences between calculation software in terms of morphologies and types of phantoms used have an impact on FD calculations? Findings Software choice has an impact on calculated FD, but is not expected to alter patient management except for extreme cases with multiple CT exams. Clinical relevance The FD limit of 100 mGy, defined by the International Commission on Radiological Protection, cannot be reached with a single CT examination, and may only be of concern in cases where the patient undergoes multiple exams with the whole fetus exposed.
More Related Videos
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
08:41Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT