Performance of Continuous Digital Monitoring of Vital Signs with a Wearable Sensor in Acute Hospital Settings
Meera Joshi1, Fahad M Iqbal1, Mansour Sharabiani2
1Department of Surgery and Cancer, Imperial College London, London W2 1NY, UK.
Background:
Continuous vital sign monitoring using wearable sensors has gained traction for the early detection of patient deterioration, particularly with the advent of virtual wards.
Objective:
The objective was to evaluate the reliability of a wearable sensor for monitoring heart rate (HR), respiratory rate (RR), and temperature in acutely unwell hospital patients and to identify the optimal time window for alert generation.
Methods:
A prospective cohort study recruited 500 patients in a single hospital. Sensor readings were compared to standard intermittent nurse observations using Bland-Altman plots to assess the limits of agreement.
Results:
HR demonstrated good agreement with nurse observations (intraclass correlation coefficient [ICC] = 0.66, r = 0.86, p < 0.001), with a mean difference of 3.63 bpm (95% LoA: -10.87 to 18.14 bpm). RR exhibited weaker agreement (ICC = 0.20, r = 0.18, p < 0.001), with a mean difference of -2.72 breaths per minute (95% LoA: -10.91 to 5.47 bpm). Temperature showed poor to fair agreement (ICC = 0.30, r = 0.39, p < 0.001), with a mean difference of -0.57 °C (95% LoA: -1.72 to 0.58 °C). A 10 min averaging window was identified as optimal, balancing data retention and real-time alerting.
Conclusions:
Wearable sensors demonstrate potential for reliable continuous monitoring of vital signs, supporting their future integration into real-world clinical practice for improved patient safety.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Introduction to Vital Signs
Vital signs help healthcare professionals assess an individual's well-being and detect any functional changes...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...