Analysis of Volatile and Oxidation Sensitive Compounds Using a Cold Inlet System and Electron Impact Mass Spectrometry
Published on: September 5, 2014
Ion mobility spectrometer with heated sample inlet - Solution to the issue of temperature effect on resolving power
Martin Lippmann1, Moritz Hitzemann1, Timo Sawatzki1
1Leibniz Universität Hannover, Institute of Electrical Engineering and Measurement Technology, Department of Sensors and Measurement Technology, Appelstr. 9A, 30167, Hannover, Germany.
Background:
One major challenge in detecting less volatile compounds with an ion mobility spectrometer (IMS) is preventing condensation of target molecules in the sampling line and ionization region to allow for fast response and recovery. Heating the entire device including the sampling line can of course mitigate condensation of such compounds, but this comes at the cost of reduced resolving power and compromised detection limits. Furthermore, a considerable amount of additional power and an IMS design with temperature-resistant components are needed.
Results:
In this work, a different approach has been investigated, with a heated sample inlet in combination with a directed sample gas flow through the ionization region, but with the drift region at lower temperature. While this approach effectively addresses the issue of condensation, it results in an inhomogeneous temperature distribution within the drift region. Simulations and experimental data reveal that this uneven temperature distribution can significantly distort the peaks in the ion mobility spectrum, depending on the IMS orientation. However, positioning the IMS vertically, with the detector facing down, significantly minimizes temperature-induced peak distortion, thereby maintaining high resolving power. In this orientation, the IMS used in this work shows a resolving power of 80, while the IMS sample inlet and outlet are heated to 423 K. In addition, a directed sample gas flow in the ionization region is used to further reduce condensation in the ionization region.
Significance And Novelty:
The approach and findings revealed in this work allow the construction of IMS with a heated sample inlet to prevent condensation of less volatile compounds while also maintaining the high resolving power of a drift tube at room temperature. The results on the influence of the orientation of the IMS can also be applied to even higher temperatures.
More Related Videos
Related Concept Videos
Atomic Spectroscopy: Effects of Temperature
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Capillary Electrophoresis: Instrumentation
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
IR Spectrometers