Roche Digital Pathology Dx whole slide imaging system is comparable to traditional microscopy for primary diagnosis
Keith A Wharton1, Jim Ranger-Moore1, Hon Seng1
1Roche Diagnostics Solutions, Oro Valley, AZ, United States.
Objective:
We evaluated the clinical performance of Roche Digital Pathology Dx, a whole slide imaging (WSI) system, in 2 studies according to US Food and Drug Administration (FDA) and Digital Pathology Association criteria.
Methods:
Precision was measured by pathologists identifying 23 histopathology features; accuracy was assessed by comparing diagnoses from 2047 clinical cases with those from manual microscopy, with exploratory analyses including subgroup-specific diagnostic discrepancy rates.
Results:
Both studies met all predetermined primary endpoints. Precision between systems/sites was 89.3%; between days, 90.3%; and between readers, 90.1% (lower bound of 95% CI for each, ≥85%). The difference in accuracy between digital reads (DRs) and manual microscopy reads (MRs) vs reference sign-out diagnosis (SD), DRs - MRs, was -0.61% (lower bound of 95% CI, -1.59%), which was greater than the lower bound acceptance criterion (-4%). Mean case reading times were similar: 2.33 minutes (DRs) and 2.34 minutes (MRs). Review of breast, lung, bladder, kidney, and stomach case diagnoses did not identify DR modality-specific root causes for major diagnostic disagreements. Higher than expected disagreements in both modalities were traced to COVID-19 pandemic-related resource constraints, leading to challenging case adjudications and higher disagreement rates for longer SDs. Direct DR/MR adjudication supported this hypothesis, resulting in an intermodality disagreement rate of 4.77%; using SD as a "tiebreaker" reduced the overall DR disagreement rate to 2.97%.
Conclusions:
Roche Digital Pathology Dx is noninferior to manual microscopy for primary diagnosis in surgical pathology, with performance results similar to 5 distinct FDA-cleared WSI systems using different scanners.
Related Concept Videos
Three-Dimensional Microscopy in Microbiology
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...