Related Experiment Video
Updated: May 2, 2026

Insertion, Maintenance, and Removal of the Percutaneous Dual Lumen Cannula Right Ventricular Assist Device
Published on: July 20, 2022
Heterotopic Crosscaval Transcatheter Tricuspid Valve Replacement for Patients With Tricuspid Regurgitation: The
Philipp Lurz1, Karl-Patrik Kresoja1, Christian Besler2
1Department of Cardiology, Cardiology I, University Medical Center of the Johannes Gutenberg-University Mainz, Mainz, Germany; German Center for Cardiovascular Research, partner site Rhine-Main, Mainz, Germany.
Background:
Transcatheter tricuspid interventions are an emerging therapy for patients with severe tricuspid regurgitation (TR), but a significant proportion of patients with TR are unsuitable for orthotopic tricuspid valve (TV) approaches.
Objectives:
The aim of this study was to investigate the performance of a dedicated heterotopic valve replacement system in patients with severe or greater TR.
Methods:
This single-arm, multicenter study assessed the technical performance, efficacy, and safety of heterotopic cross-caval transcatheter TV replacement using the Trillium device at 30 days. Patients with severe or greater TR ineligible for surgical or other interventional orthotopic TV treatment were included.
Results:
Between 2021 and 2023, 20 patients (median age 81 years, 40% women) were included. Device implantation was successful in all cases, and no intraprocedural deaths or conversions to surgery occurred. Postprocedurally, central venous pressure was reduced (median -3 mm Hg; Q1-Q3: -5 to -1 mm Hg; P = 0.003), and TR severity as assessed at the level of the Trillium device was reduced in all cases (P < 0.001). Within the 30-day follow-up period, 1 patient died, 2 patients experienced heart failure hospitalization, 2 required hemodialysis, and 2 experienced major bleeding. There was an improvement in NYHA functional class (P = 0.005) and a trend toward a reduction in edema severity score (median -0.5; Q1-Q3: -1 to -0; P = 0.052).
Conclusions:
The Trillium device is a safe and effective device for the treatment of patients with severe or greater TR ineligible for other therapeutic options. The device leads to a hemodynamic and symptomatic improvement of the burden of TR, whether this translates into favorable outcomes is to be proven in randomized trials.
More Related Videos
09:57Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Related Concept Videos
Mitral Regurgitation III: Medical Management
Mitral Stenosis III: Medical Management