Related Experiment Video
Updated: Sep 19, 2025

Procedures of Laboratory Fumigation for Pest Control with Nitric Oxide Gas
Published on: November 24, 2017
Nitrogen Dioxide Sterilization Follows Log-Linear Microbial Inactivation Kinetics Using Geobacillus
Thomas P Richards1, Delaney Lisco2, Tiffany Bianchi2
1IM3 Inc., Washougal, WA; and trichards@im3consultinggroup.com.
Abstract:
The primary purpose of this study was to determine the inactivation kinetics of Geobacillus stearothermophilus biological indicators (BIs) exposed to nitrogen dioxide (NO2) gas in the presence of humidity. BIs inoculated with 6 log10 G. stearothermophilus spores were used as a test substrate. Three BI lots manufactured from each of three different BI spore crops were evaluated. Test cycles were run at room temperature with approximately 80% relative humidity. Direct enumeration methods were used to quantify the resistance of spores with surviving populations greater than approximately 50 colony forming units (CFU). Fraction negative methods were used to estimate the surviving spore populations in the quantal region. The methods were combined in order to show spore inactivation from 6 log10 to approximately -2 log10 The D value and coefficient of determination (r2) were calculated. Over 100 direct enumeration and fraction negative cycles were completed at a fixed NO2 concentration varying only exposure time. Process parameters were maintained over all cycles. Empirical data confirmed a log-linear relationship over an 8 log10 population range with r2 values greater than 0.8, allowing for extrapolation of the curve to achieve a sterility assurance level (SAL) of 10-6 Study outcomes were comparable for all manufactured BI lots. NO2 sterilization follows first-order log-linear microbial inactivation kinetics, which is consistent with a mechanism of action based on a single active species. This is the first study to report on the microbial inactivation kinetics of NO2 sterilization.
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Physical Methods for Controlling Microbial Growth: Temperature
Biological Methods for Microbial Control
Methods for Controlling Microbial Growth