Multi-Institutional Survey of 4-Year Ultrasound Quality Control Findings
Zaiyang Long1, Megan Russ2, Sandra Larson3
1Department of Radiology, Mayo Clinic, Rochester, MN, USA.
Objectives:
Periodic quality control (QC) testing of ultrasound (US) imaging systems is essential to ensure and maintain image quality and safety. The study aims to analyze QC findings from medical physics annual surveys of modern clinical US systems in a multi-institutional survey.
Methods:
QC results from annual surveys between 2018 and 2021 were retrospectively collected from 12 medical physicists from 11 institutions or consulting companies (hereafter referred to as sites). QC tests were classified as scanner- and transducer-related tests, where scanner-related tests also included evaluation of diagnostic workstation display monitors and routine QC programs if applicable. Test methodology and pass/fail criteria were established by each site. QC findings were defined as results requiring follow-up action by service engineers, vendors, or other personnel. The percentage of annual findings was calculated for each site and averaged across the years of data contributed. Furthermore, findings across all sites were aggregated for an overall analysis.
Results:
QC data from a total of 1069 scanner tests and 4542 transducer tests were collected. The average annual percentage of scanner-related findings varied from 0 to 40.8% among all sites (first quartile 8.7%, median 18.2%, and third quartile 34.1%), while that of transducer-related findings ranged between 0.5% and 19.9% (first quartile 3.9%, median 7.2%, and third quartile 14.5%). For scanner-related tests, the top 3 categories of findings were associated with physical and mechanical integrity (171 findings, 74.7%), quantitative testing of the scanner display monitor (40 findings, 17.5%), and scanner port creating artifacts (10 findings, 4.4%). The top 3 findings for transducers related to uniformity and artifact (227 findings, 79.9%), physical and mechanical integrity (44 findings, 15.5%), and sensitivity in fundamental mode (9 findings, 3.2%).
Conclusions:
Medical physics annual surveys revealed considerable actionable findings of modern US systems, which inform the status of current US QC practices in the United States and may guide future standardization and recommendation of QC tests.
More Related Videos
10:38Observational Study Protocol for Repeated Clinical Examination and Critical Care Ultrasonography Within the Simple Intensive Care Studies
Published on: January 16, 2019
03:47Author Spotlight: Workflow for Integrating POCUS Data into EHR for Managing Heart Failure Patients
Published on: July 12, 2024
Related Concept Videos
Imaging Studies II: Ultrasonography
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: