Issues in radiation dose measurement using electronic personal dosimeters during disaster relief activities
Akira Suzuki1, Yoshiaki Hirofuji2, Noriaki Miyaji2
1Department of Radiology, Japanese Red Cross Sendai Hospital, 2-43-3 Yagiyama Honcho, Sendai, Miyagi, 982-8501, Japan. hokumeiryou@hotmail.co.jp.
Electronic personal dosimeters (EPDs) detected unusually high radiation doses in medical relief teams during the Noto Peninsula earthquake. Electromagnetic interference (EMI) from digital devices is suspected as the cause, necessitating countermeasures for radiation safety.
Area of Science:
- Medical physics
- Disaster medicine
- Radiation safety
Background:
- The Noto Peninsula earthquake necessitated medical relief efforts by the Japanese Red Cross Society (JRCS).
- Accurate radiation exposure monitoring is crucial for personnel operating in disaster zones, especially during potential nuclear incidents.
Purpose of the Study:
- To measure radiation exposure of JRCS medical relief team members during the Noto Peninsula earthquake.
- To investigate the occurrence of electromagnetic interference (EMI) events causing abnormally high dose readings on electronic personal dosimeters (EPDs).
Main Methods:
- Six JRCS medical relief team members were equipped with EPDs to monitor radiation exposure.
- A background radiation dosimeter was deployed on-site to record ambient radiation levels.
- Data collected over 2.5 days of relief activities were analyzed.
Main Results:
- Background radiation levels were minimal (3.0 ± 3.0 μSv).
- One nurse recorded an exceptionally high dose of 2075.0 ± 207.5 μSv, significantly exceeding the average dose of 40.0 ± 39.5 μSv for other team members.
- Abnormally high dose readings were observed without evidence of radioactive material dispersion.
Conclusions:
- Electromagnetic interference (EMI) from digital devices is strongly suspected as the cause of the anomalous high radiation dose readings.
- Three out of four individuals with high readings were operating digital devices, supporting the EMI hypothesis.
- Implementing countermeasures against EMI is essential for effective radiation dose management using EPDs during nuclear disaster relief operations.
More Related Videos
08:23An Automated Microscopic Scoring Method for the γ-H2AX Foci Assay in Human Peripheral Blood Lymphocytes
Published on: December 25, 2021
06:21Diffuse Optical Spectroscopy for the Quantitative Assessment of Acute Ionizing Radiation Induced Skin Toxicity Using a Mouse Model
Published on: May 27, 2016
Related Concept Videos
Electronic Distance Measuring Instruments
Biological Effects of Radiation