Published on: September 7, 2018
An enhanced-performance multisensing progressive cellular μGC: design advances and blind test results
Declan Winship1, Weilin Liao1, Hsueh-Tsung Lu2
1Department of Electrical Engineering and Computer Science, and Center for Wireless Integrated MicroSensing and Systems (WIMS²), University of Michigan, Ann Arbor, MI, 48109, USA.
This study presents advancements in microscale gas chromatographs (µGCs) for reliable in-field chemical vapor analysis. Key improvements include enhanced humidity tolerance, precise flow control for repeatable results, and an onboard reference standard for long-term deployment.
Area of Science:
- Analytical Chemistry
- Chemical Engineering
- Sensor Technology
Background:
- In-field chemical vapor analysis is crucial for environmental, industrial, and security applications.
- Microscale gas chromatographs (µGCs) offer potential but require improvements in repeatability, humidity tolerance, and in-field calibration.
- Existing µGCs face challenges with environmental factors and drift, limiting their reliability for continuous monitoring.
Purpose of the Study:
- To report chip-level and system-level advancements in µGC technology for reliable in-field chemical analysis.
- To enhance the performance of µGCs regarding thermal management, humidity response, retention time repeatability, and calibration.
- To demonstrate the practical utility of the improved µGC for chemical screening and long-term field deployment.
Main Methods:
- Utilized a µGC with monolithic integration of preconcentrators, separation columns, and capacitive/photoionization detectors.
- Implemented tailored heater designs for advanced thermal management and incorporated fence electrodes in photoionization detectors to mitigate humidity effects.
- Introduced closed-loop flow control for improved retention time repeatability and integrated a miniature onboard chemical reference standard reservoir.
Main Results:
- Humidity response in photoionization detectors was reduced by over 98% using fence electrodes.
- Closed-loop flow control improved retention time repeatability, achieving a relative standard deviation of 0.29-0.43% (4-5× improvement over open-loop).
- Successful blind tests for false alarm and chemical recognition demonstrated the instrument's capability for accurate analyte identification and screening.
Conclusions:
- The developed µGC instrument with integrated modules shows significant promise for reliable, long-term in-field chemical analysis.
- Advancements in thermal management, humidity compensation, flow control, and onboard calibration address key limitations of previous µGC systems.
- The instrument is well-suited for broad chemical screening applications requiring high repeatability and robustness in diverse environmental conditions.