Published on: May 30, 2019
Accuracy of current intraoral scanning systems for full-arch impressions: An in vitro study
Jenny Buhl1, Andreas Ender1, Albert Christof Mehl1
1Division of Computerized Restorative Dentistry, Clinic for Preventive Dentistry, Periodontology and Cariology, Center for Dental Medicine, University of Zurich, Plattenstrasse 11, 8032 Zurich, Switzerland.
Objective:
This in vitro study aimed to evaluate the accuracy of seven intraoral scanners (IOSs) for full-arch impressions. Correlations among three evaluation parameters summarising three-dimensional (3D) difference values were also analysed.
Methods:
Custom-made complete-arch reference models were initially scanned using a reference scanner (ATOS III, Triple Scan MV60). Full-arch scans were then performed using seven IOSs (n = 20 each): Aoralscan 3, Emerald Green (EME), iTero 5D, Medit i900, Primescan (PRI), Primescan 2, and Trios 5. Conventional impressions (CONVs) (n = 20) were obtained and digitised using a laboratory scanner (inEOS X5). All data were exported as standard tessellation language files. Trueness and precision were determined using (90_10)/2-percentile values. Statistical analyses were performed using analysis of variance with Games-Howell post-hoc test and linear mixed-effects model. Correlations among absolute 80-percentile, absolute mean, and (90_10)/2-percentile were analysed using linear regression with Pearson's correlation coefficient.
Results:
Trueness ranged from 24.5 ± 3.2 μm (CONV) to 95.0 ± 18.8 μm (EME), while precision ranged from 13.3 ± 2.6 μm (CONV) to 59.6 ± 22.4 μm (EME). CONV outperformed all IOSs. PRI showed the highest accuracy (trueness: 30.5 ± 5.3 μm, precision: 18.7 ± 5.6 μm). The evaluated parameters showed highly significant linear correlations (p < 0.001).
Conclusions:
All tested IOSs show full-arch deviations of <100 µm. The evaluation parameters are highly correlated, allowing for comparability.
Clinical Significance:
IOSs reach clinically acceptable accuracy values for full-arch impressions. However, high-precision CONVs still achieve the highest trueness and precision for full-arch impressions.