Safire CT iterative reconstruction protocol to reduce patient effective dose on PET/CT examinations
Alice Ferretti1, Sotirios Chondrogiannis2, Jlenia Peretto2
1Medical Physics Unit, Santa Maria della Misericordia Hospital of Rovigo, ULSS5 Polesana, Rovigo, Italy.
Introduction:
To implement the use of CT iterative reconstruction, instead of standard FBP, in order to reduce the patient dose of the 18F-FDG PET/CT examinations, preserving image quality and SUV accuracy.
Methods:
Catphan 600 and Alderson-Rando anthropomorphic phantom were used to evaluate the SAFIRE iterative CT reconstruction on Siemens Biograph mCT system, progressively reducing theCareDose4D Quality-Ref-mAs. The PET quantitative accuracy was also evaluated by NEMA2012 IQ phantom. The optimized CT reconstruction protocol was applied in clinical setting on additional segmental PET/CT scan, after standard whole body PET/CT, for qualitative evaluation by two nuclear medicine physicians.
Results:
Measurements on Cathpan phantom demonstrated thatSafire reconstruction level 3 allowed to reduce the CT dose of about 40% without worsening the noise and CNR. For mAs-modulated acquisitions on the anthropomorphic phantom, a 30% reduction of doseusingSafire at level 3does not worsen image quality in any anatomical part. Analysing PET/CT acquisitions of NEMA IQ phantom with standard CT FBP reconstruction vs dose-optimized CT iterative reconstruction, the SUV values showed very small deviations (<0.6 %). The clinical evaluation carried out on 25 patients demonstrates the non-inferiority of the image quality obtained with the reduced CT dose and Safire 3 iterative reconstruction protocol compared to the reference CT dose and FBP reconstruction.
Conclusions:
The use of the Safire iterative reconstruction at level 3 in whole-body PET/CT acquisition allowed an effective dose reduction of 1.8 mSv in mean (from 10.1 mSv to 8.3 mSv, p < 0.001), without any detriment of image quality and SUV accuracy.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET