Related Experiment Video
Updated: Sep 13, 2025

Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022
'Inverted Tulip': New Deformation of Atrial Septal Defect Closure Devices
Kevin Velarde-Acosta1, Josh Yefry Moscoso Ramirez1, Paol Jhullian Rojas1,2
1Clinical Cardiology Department, Hospital Guillermo Almenara Irigoyen - EsSalud Lima, Peru.
Abstract:
We present the case of a 45-year-old woman (62 kg, 160 cm) who suffered an ischemic stroke associated with patent foramen ovale (PFO) with high-risk anatomical features. The defect was percutaneously closed using a LifeTech CeraFlex PFO 30 mm × 25 mm device deployed via a 12 Fr sheath after multiple implantation attempts. Following release, the right atrial disc exhibited a concave deformation, termed the 'inverted tulip,' which prolonged the procedure and resulted in a significant residual shunt. The shunt resolved spontaneously at the 1-month echocardiographic follow-up.