Rapid Molecular Detection and Differentiation of Influenza Viruses A and B
Published on: January 30, 2017
Diagnostic accuracy evaluation of a point-of-care antigen test for SARS-CoV-2 and influenza in UK primary care
Thomas R Fanshawe1, Sharon Tonner1, Philip J Turner1
1Nuffield Department of Primary Care Health Sciences, University of Oxford, Oxford, United Kingdom.
Objectives:
To evaluate the diagnostic accuracy of the Roche SARS-CoV-2 & Flu A/B Rapid Antigen Test at the point of care.
Design:
Prospective diagnostic accuracy study.
Setting:
17 primary care practices in England.
Participants:
500 individuals with symptoms consistent with possible SARS-CoV-2 or influenza infection identified upon presentation to primary care or via medical note review.
Primary And Secondary Outcome Measures:
Sensitivity, specificity and predictive values, compared to a laboratory reference standard of real-time reverse transcription PCR, using samples collected using a combined nasal and oropharyngeal swab.
Results:
Of 481 participants with available index and reference test results, 5.6% (27/481) were reference standard-positive for SARS-CoV-2, 11.4% (55/481) for Influenza A and 1.9% (9/481) for Influenza B. The sensitivity of the antigen test to detect SARS-CoV-2 was 70.4% (19/27, 95% CI 49.6-86.2%) and specificity was 99.3% (451/454, 95%CI 98.1-99.9%). For Influenza A, sensitivity was 29.1% (16/55, 95% CI 17.6-42.9%) and specificity 98.6% (420/426, 97.0-99.5%), and for Influenza B, sensitivity was 22.2% (2/9, 2.8-60.0%) and specificity 98.1% (463/472, 96.4-99.1%).
Conclusions:
In a primary care population of symptomatic individuals, the assay was highly specific and had moderate sensitivity to detect SARS-CoV-2, but did not detect the majority of influenza infections.
Registration:
ISRCTN14226970.