Speckle contrast optical spectroscopy for cuffless blood pressure estimation based on microvascular blood flow and
Ariane Garrett1, Byungchan Kim1, Nil Z Gurel2
1Department of Biomedical Engineering, Boston University, Boston, MA 02215, USA.
Abstract:
This work introduces high-speed (390 Hz) speckle contrast optical spectroscopy (SCOS) to enable simultaneous measurements of multi-anatomic site microvascular blood volume and flow oscillations. Simultaneous blood flow and volume waveforms were extracted at two wavelengths on the wrist and finger, in reflectance and transmission mode, respectively. Blood volume changes (also known as photoplethysmography, or PPG) were determined based on intensity oscillations. Blood flow information was determined based on dynamic light scattering information encoded in the 2D spatial speckle pattern after removal of stochastic and instrument noise. We extracted a wide array of temporal, shape-based, and frequency-domain features from each high-resolution waveform, as well as features that characterize the temporal relationships between these features. These features and their inter-relationships are determined by the dynamic biomechanical properties of peripheral microvasculature, including vascular compliance and resistance, which are key determinants of dynamic changes in systemic blood pressure (BP). In comparison to PPG alone, SCOS demonstrated a notable 31% improvement (p = 3.45 * 10-7) in systolic BP estimation when integrated into subject-specific machine-learning models. The resulting errors were remarkably low (systolic BP: 0.06+/- 2.88 mmHg, diastolic BP: 0.09 +/-2.14 mmHg) across a wide range of BP variations (range SBP: 89-284 mmHg). This improvement was sustained several weeks later within a re-measured cohort, indicating highly robust BP predictions. Looking ahead, high-speed SCOS holds the potential to substantially enhance the non-invasive characterization of the cardiovascular system, including continuous and non-invasive BP measurements, which are a long-sought-after goal of the biomedical community.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
05:24How to Build a Laser Speckle Contrast Imaging LSCI System to Monitor Blood Flow
Published on: November 11, 2010
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Korotkoff Sounds
During blood pressure assessment, inflating the cuff 30 millimeters of mercury above the patient's systolic blood pressure...