Photon-counting CT versus energy-integrating detector and flat-panel CT for cadaveric wrist arthrography with
Johannes de Boer1, Nigar Salimova2, Friederike Weidemann3
1Department of Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany. johannes.dbh@gmail.com.
Background:
This study aimed to evaluate the imaging performance and diagnostic value of a photon-counting detector (PCD) computed tomography (CT) compared to an energy-integrating detector (EID) and flat panel detector (FPD) for cadaveric wrist arthrographies.
Methods:
Following ethics committee approval, ten cadaveric wrists were injected with diluted iodinated contrast agent. CT arthrographies using PCD-, EID-, and FPD-CT were performed. Six dose protocols between 0.1 mGy (using a tin filter) and 6 mGy, ultrahigh-resolution-mode, and two reconstruction kernels were used for the PCD-CT and EID-CT. FPD-CT images were reconstructed using a "normal" and "sharp" kernel. Signal-to-noise ratios (SNR) and contrast-to-noise ratios (CNR) were calculated and analyzed using analysis of variance (ANOVA) and post hoc tests. Three blinded radiologists independently rated image quality concerning trabecular, cartilage, and intrinsic structures. Intraclass correlation coefficients (ICC) were calculated, followed by a Friedman and post hoc test.
Results:
At 1.5 mGy, 3 mGy, and 6 mGy with the Br89 kernel, the PCD-CT yielded up to 2.35 times higher SNR and up to 7 times higher CNR than dose-equivalent and higher dose EID-CT scans. Subjective ratings favored the PCD-CT over the EID-CT and occasionally the FPD-CT, with a combined ICC of 0.942. Applying sharper kernels, SNR did not differ significantly between the PCD-CT (1.5 mGy, 3 mGy, and 6 mGy) and the FPD-CT.
Conclusion:
Using sharp kernels, the PCD-CT provided superior image quality to the EID-CT and achieved comparable or better quality than the FPD at certain parameters. Thus, the PCD-CT could be considered a possible alternative in clinical routine for evaluating wrist injuries.
Relevance Statement:
This study demonstrates the potential of the PCD-CT as a valuable tool in diagnosing wrist injuries. Its superior image quality compared to the EID-CT can increase confidence in diagnosing subtle bone pathologies and additionally yields the possibility of radiation exposure reduction.
Key Points:
The technical advantages of the PCD-CT allow for dose reduction while generating high-quality images. PCD-CT showed superior image quality over EID-CT and was comparable to the FPD-CT. PDC-CT offers improved visualization of fine joint structures in wrist arthrography and should be considered in clinical routine.
More Related Videos
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT